Repeat patient testing-based quality control shows promise for use in veterinary biochemistry testing
Bente Flatland1, Kathleen P Freeman1,2
1Department of Biomedical and Diagnostic Sciences, College of Veterinary Medicine A205, University of Tennessee, Knoxville, TN, USA.
Veterinary Clinical Pathology
|December 14, 2020
Summary
Canine heparinized plasma is suitable for repeat patient testing-based quality control (RPT-QC), offering comparable analytical error detection to commercial controls. This method is viable for both reference and in-clinic laboratories.
Area of Science:
- Clinical Chemistry
- Laboratory Quality Control
- Veterinary Diagnostics
Background:
- Repeat patient testing-based quality control (RPT-QC) offers a statistical alternative to traditional commercial quality control materials (QCM).
- Evaluating novel biological matrices for RPT-QC is crucial for expanding its application.
- Canine heparinized plasma presents a potential matrix for RPT-QC due to its availability and characteristics.
Purpose of the Study:
- To assess the suitability of canine heparinized plasma for RPT-QC.
- To evaluate the performance of RPT-QC in detecting analytical errors in clinical chemistry testing.
- To compare the efficacy of RPT-QC with canine plasma against QCM-QC.
Main Methods:
- Investigated canine plasma stability under refrigerated conditions (7 days) at varying time points.
- Validated RPT-QC limits for 17 common analytes using prospectively measured plasma.
- Challenged established RPT-QC limits by introducing simulated analytical errors.
- Compared RPT-QC performance with QCM-QC using Westgard's EZ Rules 3 for four control rules.
Main Results:
- Canine plasma demonstrated acceptable stability for RPT-QC after 7 days of refrigeration.
- RPT-QC limits were successfully validated for 12 out of 17 common measurands.
- Validated RPT-QC effectively identified simulated analytical errors.
- RPT-QC demonstrated performance comparable to QCM-QC in detecting analytical errors across common control rules.
Conclusions:
- Canine heparinized plasma, when refrigerated for 7 days, is a suitable matrix for RPT-QC.
- RPT-QC with canine plasma can be implemented with simple control rules and minimal control materials.
- The findings support the applicability of canine plasma-based RPT-QC in both reference and in-clinic laboratory settings.