Related Experiment Video
Updated: Nov 26, 2025

Upper-extremity Approach for Secondary Access in Transfemoral Transcatheter Aortic Valve Implantation
Published on: August 8, 2025
Self-expanding transcatheter aortic valve infolding: Current evidence, diagnosis, and management
Marco Bruno Ancona1, Alessandro Beneduce1, Vittorio Romano1
1Interventional Cardiology Unit, IRCCS San Raffaele Scientific Institute, Milan, Italy.
Background:
Prosthetic valve infolding is a rare but severe complication of transcatheter aortic valve implantation (TAVI) with self-expanding valves. However, currently available clinical data are limited and fragmented.
Objectives:
This report aims to provide a comprehensive overview of this complication focusing on predisposing factors, clinical presentation, diagnostic findings, treatment and clinical outcomes.
Methods:
A systematic review of the literature was performed to identify cases of infolding occurring during TAVI with self-expanding valves published until August 2020. These data were pooled with all the retrospectively identified infolding cases occurred at San Raffaele Scientific Institute between December 2014 and August 2020.
Results:
A total of 34 cases were included. Among patients with available data, 38% received a first-generation CoreValve, and 62% a second-generation Evolut R (82%) or Evolut PRO (18%). Infolding occurred mostly with ≥29-mm valves (94%). Predisposing factors included resheathing of a second-generation valve (82%), heavy calcification of the native valve (65%), lack of predilatation (16%), Sievers type-1 bicuspid aortic valve (11%), and improper valve loading (5%). Infolding resulted in severe PVL causing hemodynamic instability (29%) or cardiac arrest (12%). Postdilatation was the treatment strategy in 68%, while prosthesis replacement with a new device in 23% of cases. Device success rate was 82%. Death and stroke occurred in 3% and 12% of cases.
Conclusions:
Prosthetic valve infolding is typically observed after resheathing of a large-size self-expanding TAVI. When infolding is timely diagnosed, prosthesis removal and replacement should be pursued. Further studies are required to precisely define predisposing factors to prevent this complication.
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
08:50Technique and Patient Selection Criteria of Right Anterior Mini-Thoracotomy for Minimal Access Aortic Valve Replacement
Published on: March 26, 2018
Related Concept Videos
Aortic Regurgitation III: Medical Management
Mitral Valve Prolapse II: Assessment and Management
Aortic Regurgitation I: Introduction
Mitral Stenosis III: Medical Management
Aortic Regurgitation IV: Nursing Management
Mitral Valve Prolapse III: Nursing Management