Related Experiment Video
Updated: Nov 26, 2025

Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Wavefront-guided myopic laser in situ keratomileusis with a high-resolution Hartmann-Shack aberrometer and a new
Robert K Maloney1, Colman R Kraff, Stephen C Coleman
1From the Maloney-Shamie Vision Institute, Los Angeles, California (Maloney); Kraff Eye Institute, Chicago, Illinois (Kraff); Coleman Vision, Albuquerque, New Mexico (Coleman).
Purpose:
To evaluate visual, refractive, aberrometric, and patient-reported outcomes of wavefront-guided (WFG) myopic laser in situ keratomileusis (LASIK) using a high-resolution Hartmann-Shack aberrometer (iDesign Advanced WaveScan system) with a new nomogram and to determine whether the new nomogram resolved the mild undercorrection that occurs with the manufacturer's default settings.
Setting:
Three private LASIK practices.
Design:
Prospective, open-label, noncomparative, multicenter study.
Methods:
One hundred ninety eyes of 95 patients underwent bilateral WFG LASIK for the correction of myopia or myopic astigmatism. A new nomogram was used, which effectively adjusted the wavefront-measured refraction sphere up or down to equal the manifest refraction sphere. Patients were followed up for 6 months.
Results:
Eighty-four patients completed the final follow up. At 6 months, 162 (96.4%) of 168 eyes achieved monocular uncorrected distance visual acuity of 20/20 or better. No eye lost 2 or more lines of corrected distance visual acuity. The safety and efficacy indices were 1.12 and 1.09, respectively; 164 (98%) of 168 eyes had manifest refraction spherical equivalent within ±0.50 diopters (D) of emmetropia, and 154 (92%) of 168 eyes had residual manifest refractive astigmatism of 0.50 D or less. Fewer patients experienced burning, stinging, soreness, and irritation postoperatively than preoperatively. Eighty-one (96%) of 84 patients reported improved quality of life.
Conclusions:
WFG myopic LASIK using a high-resolution Hartmann-Shack aberrometer and a new nomogram resolved the undercorrection with the manufacturer's default settings. The treatment was safe and effective with excellent visual and refractive outcomes, high patient satisfaction, and improved quality of life.
More Related Videos
Related Concept Videos
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...