Related Experiment Video
Updated: Nov 25, 2025

An Application for Pairing with Wearable Devices to Monitor Personal Health Status
Published on: February 3, 2022
iPhone App compared with standard blood pressure measurement -The iPARR trial
Marcus Dörr1, Stefan Weber2, Ralf Birkemeyer3
1Department of Internal Medicine, University Hospital Greifswald, Germany; DZHK (German Centre for Cardiovascular Research), partner site Greifswald, Germany.
Background:
The possibility to use built-in smartphone-cameras for photoplethysmographic (PPG) recording of pulse waves lead to the release of numerous health apps, claiming to measure blood pressure (BP) based on PPG signals. Even though these apps are highly popular, not a single one is clinically validated. Aim of the current study was to test systolic BP (sBP) estimation by a promising new algorithm in a large clinical setting.
Methods:
The study was designed based on the European Society of Hypertension International Protocol Revision 2010. Each individual received 7 sequential BP measurements, starting with the reference device - an automated oscillometric cuff device - followed by the PPG recording at the patients' index finger.
Results:
A total 1,036 subjects were recruited of which 965 could be included for final analysis leading to 2,895 pairs of comparison. Mean (±SD) error between test and reference device was -0.41 (±16.52) mmHg. Only 38.1% of all 2,895 BP comparisons reached a delta within ±5 mmHg, while 29.3% reached a delta larger than 15 mmHg. Bland-Altman plot showed an overestimation of smartphone sBP in comparison to reference sBP in low range and an underestimation in high sBP range.
Conclusions:
According to the European Society of Hypertension International Protocol Revision 2010 specifications the algorithm failed validation criteria for sBP measurement and was not commercialized. These findings emphasize that health apps should be rigorously validated according to common guidelines before market release as under- and/or overestimation of BP is potentially exposing persons at health risks in short and long term.
Trial Registration:
ClinicalTrials.gov, number NCT02552030.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure