First Outpatient Evaluation of a Tubeless Automated Insulin Delivery System with Customizable Glucose Targets in
Gregory P Forlenza1, Bruce A Buckingham2, Sue A Brown3
1Barbara Davis Center for Diabetes, University of Colorado School of Medicine, Aurora, Colorado, USA.
Insights
The Omnipod 5 system improved time in range for children and adults with type 1 diabetes (T1D). This tubeless automated insulin delivery system demonstrated safety and effectiveness in an outpatient setting.
Area of Science:
- Endocrinology
- Biomedical Engineering
- Diabetes Technology
Background:
- Type 1 diabetes (T1D) management requires continuous glucose monitoring and insulin delivery.
- Automated insulin delivery (AID) systems offer improved glycemic control and quality of life.
- The Omnipod 5 is a novel tubeless AID system utilizing hybrid closed-loop (HCL) technology.
Purpose of the Study:
- To evaluate the safety and effectiveness of the Omnipod 5 system in children and adults with T1D.
- To compare glycemic control during standard therapy (ST) versus HCL phases using the Omnipod 5.
- To assess device-related adverse events in an outpatient setting.
Main Methods:
- A single-arm, multicenter, prospective clinical study involving 36 participants (children and adults with T1D).
- A 14-day ST phase followed by a 14-day HCL phase with varying glucose targets (110-150 mg/dL).
- Primary endpoints included time in range (TIR) and incidence of serious device-related adverse events.
Main Results:
- Significantly higher mean TIR was observed during the HCL phase compared to the ST phase across different glucose targets.
- Children showed improved TIR, particularly with lower glucose targets (e.g., 110 mg/dL).
- No serious device-related adverse events, severe hypoglycemia, or diabetic ketoacidosis were reported during the HCL phase.
Conclusions:
- The Omnipod 5 system is safe and effective for managing T1D in children and adults in an outpatient setting.
- The system significantly improves time in range, indicating enhanced glycemic control.
- The tubeless AID technology provides a viable option for T1D management with remote safety monitoring.
Abstract:
The objective of this study was to assess the safety and effectiveness of the first commercial configuration of a tubeless automated insulin delivery system, Omnipod® 5, in children (6-13.9 years) and adults (14-70 years) with type 1 diabetes (T1D) in an outpatient setting. This was a single-arm, multicenter, prospective clinical study. Data were collected over a 14-day standard therapy (ST) phase followed by a 14-day hybrid closed-loop (HCL) phase, where participants (n = 36) spent 72 h at each of three prespecified glucose targets (130, 140, and 150 mg/dL, 9 days total) then 5 days with free choice of glucose targets (110-150 mg/dL) using the Omnipod 5. Remote safety monitoring alerts were enabled during the HCL phase. Primary endpoints were difference in time in range (TIR) (70-180 mg/dL) between ST and HCL phases and proportion of participants reporting serious device-related adverse events. Mean TIR was significantly higher among children in the free-choice period overall (64.9% ± 12.2%, P < 0.01) and when using a 110 mg/dL target (71.2% ± 10.2%, P < 0.01), a 130 mg/dL target (61.5% ± 7.7%, P < 0.01), and a 140 mg/dL target (64.8% ± 11.6%, P < 0.01), and among adults using a 130 mg/dL target (75.1% ± 11.6%, P < 0.05), compared to the ST phase (children: 51.0% ± 13.3% and adults: 65.6% ± 15.7%). There were no serious device-related adverse events reported during the HCL phase, nor were there episodes of severe hypoglycemia or diabetic ketoacidosis. The Omnipod 5 System was safe and effective when used at glucose targets from 110 to 150 mg/dL for 14 days at home in children and adults with T1D.
More Related Videos
08:32Studying the Hypothalamic Insulin Signal to Peripheral Glucose Intolerance with a Continuous Drug Infusion System into the Mouse Brain
Published on: January 4, 2018
08:01A Method for Manipulating Blood Glucose and Measuring Resulting Changes in Cognitive Accessibility of Target Stimuli
Published on: August 12, 2016
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Diabetes Mellitus: Type 2 and Gestational
Diabetes: Management and Pharmacotherapy
Insulin remains the cornerstone of treatment for most patients with type 1 and many...
Diabetes Mellitus: Overview and Type I Subtype
Type 1 diabetes is an autoimmune disease in which the immune system mistakenly attacks and destroys the insulin-producing beta cells in the pancreas. As a result, the body is unable to produce sufficient insulin, and individuals with...
Diabetes: Symptoms, Diagnosis, and Complications
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into...