Measuring 3D In-vivo Shoulder Kinematics using Biplanar Videoradiography
Published on: March 12, 2021
Comparative study of glenoid version and inclination using two-dimensional images from computed tomography and
Chang-Hyuk Choi1, Hee-Chan Kim1, Daewon Kang1
1Department of Orthopedic Surgery, School of Medicine, Catholic University of Daegu, Daegu, Korea.
Three-dimensional (3D) measurements of glenoid version and inclination using reconstructed bone models offer greater consistency than two-dimensional (2D) CT scans. This 3D approach provides a reliable alternative for accurate glenoid measurements.
Area of Science:
- Orthopedic Surgery
- Radiology
- Biomedical Engineering
Background:
- Accurate measurement of glenoid version and inclination is crucial for shoulder arthroplasty and diagnosing shoulder instability.
- Traditional two-dimensional (2D) computed tomography (CT) imaging may have limitations in precisely assessing these parameters.
Purpose of the Study:
- To compare the accuracy and reliability of glenoid version and inclination measurements using 2D CT images versus 3D reconstructed bone models.
- To evaluate the intra-observer and inter-observer reliability of both measurement techniques.
Main Methods:
- Thirty patients underwent conventional CT scans.
- Glenoid version and inclination were measured three times each by two independent orthopedic surgeons using 2D CT images.
- The same measurements were performed by two other surgeons using 3D reconstructed bone models generated with Mimics and 3-Matics software.
Main Results:
- Mean glenoid version was -1.705º (2D) vs. 2.635º (3D), and mean inclination was 9.08º (2D) vs. 7.23º (3D).
- Intra-observer reliability was higher for 3D measurements (0.883-0.892) compared to 2D (0.605-0.698).
- Inter-observer reliability was significantly better with 3D measurements (0.845-0.853) than 2D (0.374-0.456).
Conclusions:
- The discrepancies between 2D and 3D measurements stem from the tools used, not the image data itself.
- Three-dimensional (3D) measurements demonstrate superior intra-observer and inter-observer reliability.
- 3D measurement using reconstructed bone models is a reliable and recommended alternative for assessing glenoid version and inclination.
More Related Videos
08:15Three-Dimensional Preoperative Virtual Planning in Derotational Proximal Femoral Osteotomy
Published on: February 17, 2023
13:16Computed Tomography and Optical Imaging of Osteogenesis-angiogenesis Coupling to Assess Integration of Cranial Bone Autografts and Allografts
Published on: December 22, 2015
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...