Changes in the Plethysmographic Perfusion Index During an End-Expiratory Occlusion Detect a Positive Passive Leg
Alexandra Beurton1,2, Francesco Gavelli1, Jean-Louis Teboul1,2
1Service de médecine intensive-réanimation, Hôpital de Bicêtre, Hôpitaux universitaires Paris-Saclay, Assistance publique-Hôpitaux de Paris, Le Kremlin-Bicêtre, France.
Objectives:
The end-expiratory occlusion test for assessing preload responsiveness consists in interrupting mechanical ventilation for 15 seconds at end-expiration and measuring the cardiac index changes. The perfusion index is the ratio between the pulsatile and the nonpulsatile portions of the plethysmography signal and is, in part, determined by stroke volume. We tested whether the end-expiratory occlusion-induced changes in perfusion index could detect a positive passive leg raising test, suggesting preload responsiveness.
Design:
Observational study.
Setting:
Medical ICU.
Patients:
Thirty-one ventilated patients without atrial fibrillation.
Interventions:
We measured perfusion index (Radical-7 device; Masimo Corp., Irvine, CA) and cardiac index (PiCCO2; Pulsion Medical Systems, Feldkirchen, Germany) before and during a passive leg raising test and a 15-second end-expiratory occlusion.
Measurements And Main Results:
In 19 patients with a positive passive leg raising test (increase in cardiac index ≥ 10%), compared to the baseline value and expressed as a relative change, passive leg raising increased cardiac index and perfusion index by 17% ± 7% and 49% ± 23%, respectively, In these patients, end-expiratory occlusion increased cardiac index and perfusion index by 6% ± 2% and 11% ± 8%, respectively. In the 12 patients with a negative passive leg raising test, perfusion index did not significantly change during passive leg raising and end-expiratory occlusion. Relative changes in perfusion index and cardiac index observed during all interventions were significantly correlated (r = 0.83). An end-expiratory occlusion-induced relative increase in perfusion index greater than or equal to 2.5% ([perfusion index during end-expiratory occlusion-perfusion index at baseline]/perfusion index at baseline × 100) detected a positive passive leg raising test with an area under the receiver operating characteristic curve of 0.95 ± 0.03. This threshold is larger than the least significant change observed for perfusion index (1.62% ± 0.80%).
Conclusions:
Perfusion index could be used as a reliable surrogate of cardiac index for performing the end-expiratory occlusion test. Confirming previous results, the relative changes in perfusion index also reliably detected a positive passive leg raising test.
More Related Videos
Related Concept Videos
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...