Optimized method for in vivo dosimetry with small films in pelvic IOERT for rectal cancer
Joana Santos1, Sofia Silva2, Sandra Sarmento3
1Physics and Astronomy Department, Faculty of Sciences, University of Porto, Portugal; Medical Physics, Radiobiology and Radiation Protection Group, IPO Porto Research Center (CI-IPOP), Portuguese Oncology Institute of Porto (IPO Porto), Porto, Portugal.
Purpose:
Intra-Operative Electron Radiation Therapy (IOERT) is used to treat rectal cancer at our institution, and in vivo measurements with Gafchromic EBT3® films were introduced as quality assurance. The purpose of this work was to quantify the uncertainties associated with digitization of very small EBT3 films irradiated simultaneously, in order to optimize in vivo dosimetry for IOERT.
Methods:
Film samples of different sizes - M1 (5×5cm2), M2 (1.5×1.5 cm2), M3 (1.0×1.5 cm2) and M4 (0.75×1.5 cm2) - were used to quantify typical variations (uncertainties) due to scanner fluctuations, misalignment, film inhomogeneity, long-term effect of film cutting, small rotations, film curling, edge effects and the influence of opaque templates. Fitting functions and temporal validity of sensitometric curves were also assessed.
Results:
Film curling, intra-film variability and scanner fluctuations are important effects that need to be minimized or considered in the uncertainty budget. Small rotations, misalignments and film cutting have little or no influence on the readings. Most fitting functions perform well, but the quantity used for dose quantification determines over- or under-valuation of dose in the long term. Edge effects and the influence of opaque templates need to be well understood, to allow optimization of methodology to the intended purpose.
Conclusion:
The proposed method allows practical and simultaneous digitization of up to ten small irradiated film samples, with an experimental uncertainty of 1%.