Association between Corneal Stiffness Parameter at the First Applanation and Keratoconus Severity
Kaili Yang1, Liyan Xu1, Qi Fan1
1Henan Provincial People's Hospital, Henan Eye Hospital, Henan Eye Institute, People's Hospital of Zhengzhou University, Henan University People's Hospital, Zhengzhou 450003, China.
Corneal stiffness parameter at first applanation (SP-A1) is lower in keratoconus eyes and decreases with disease severity. This parameter can help identify keratoconus from normal eyes.
Area of Science:
- Ophthalmology
- Biomechanical Engineering
- Corneal Science
Background:
- Keratoconus is a progressive thinning disorder of the cornea.
- Accurate diagnosis and severity assessment are crucial for effective management.
- Corneal biomechanics offer potential insights into keratoconus pathophysiology.
Purpose of the Study:
- To evaluate the corneal stiffness parameter at first applanation (SP-A1) in normal and keratoconus eyes.
- To explore the association between SP-A1 and indicators of keratoconus severity.
Main Methods:
- A comparative study included 351 normal and 351 keratoconus eyes.
- Keratoconus diagnosis based on topography and slit-lamp examination.
- Corneal severity classified as mild, moderate, or severe based on steep keratometry (Ks).
- SP-A1 measured using Corvis ST software; correlation and ROC analyses performed.
Main Results:
- SP-A1 values were significantly lower in keratoconus eyes (72.11 mmHg/mm) compared to normal eyes (110.89 mmHg/mm).
- SP-A1 decreased progressively with increasing keratoconus severity (mild: 79.54, moderate: 65.11, severe: 47.59 mmHg/mm).
- SP-A1 showed a negative correlation with Ks and a positive correlation with thinnest corneal thickness (TCT).
- SP-A1 demonstrated significant associations with intraocular pressure (IOP), biomechanical IOP (bIOP), and various air-puff response parameters.
- Receiver Operating Characteristic (ROC) analysis showed high accuracy for SP-A1 in detecting keratoconus (AUC=0.952) and mild keratoconus (AUC=0.930).
Conclusions:
- SP-A1 is a valuable indicator of corneal biomechanical properties in keratoconus.
- Lower SP-A1 values correlate with increased keratoconus severity.
- SP-A1 demonstrates potential as a diagnostic tool for identifying keratoconus in clinical practice.