Radiation Dose during Transarterial Radioembolization: A Dosimetric Comparison of Cone-Beam CT and Angio-CT
Emily L Marshall1, Samuel Guajardo1, Emily Sellers1
1Department of Radiology, University of Chicago, Chicago, IL 60637.
Purpose:
To evaluate the radiation dose differences for intraprocedural computed tomography (CT) imaging between cone-beam CT and angio-CT acquired during transarterial radioembolization (TARE) therapies for hepatocellular carcinoma.
Materials And Methods:
A retrospective cohort of 22 patients who underwent 23 TARE procedures were selected. Patients were imaged in both cone-beam CT and angio-CT rooms as a part of their conventional treatment plan. Effective dose contributions from individual CT acquisitions as well as the cumulative dose contributions from procedural 3D imaging were evaluated. Angiography dose contributions were omitted. Cone-beam CT images were acquired on a C-arm Philips Allura system. Effective doses were evaluated by coupling previously published conversion factors (effective dose per dose-area product) to patient's dose-area product meter readings after the procedure. Angio-CT images were acquired on a hybrid Canon Infinix-i Aquilion PRIME system. Effective doses from angio-CT scans were estimated using Radimetrics. Comparisons of a single patient's dose differential between the 2 technologies were made.
Results:
The mean effective dose from a single CT scan was 6.42 mSv and 5.99 mSv in the cone-beam CT room and the angio-CT room, respectively (P = .3224), despite the greater field of view and average craniocaudal scan coverage in angio-CT. The mean effective dose summed across all CTs in a procedure was 12.89 mSv and 34.35 mSv in the cone-beam CT room and the angio-CT room, respectively (P = .0018).
Conclusions:
The mean effective dose per CT scan is comparable between cone-beam CT and angio-CT when considered in direct comparison for a single patient.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation I: X-ray and CT