Point-of-Care Testing Using Invasive and Non-Invasive Hemoglobinometers: Reliable and Valid Method for Estimation of
Gomathi Ramaswamy1, Kashish Vohra1, Kapil Yadav1
1Centre for Community Medicine, All India Institute of Medical Sciences, New Delhi 110029, India.
Insights
Point-of-care testing (POCT) devices like invasive digital hemoglobinometers (DHs) show promise for diagnosing anemia in children in India. While non-invasive DHs need further development, invasive DHs and Sahli
Area of Science:
- Pediatric Hematology
- Point-of-Care Diagnostics
- Public Health
Background:
- Anemia affects 58% of children aged 6-59 months in India, necessitating accurate and accessible diagnostic tools.
- Point-of-care testing (POCT) devices offer a solution for hemoglobin estimation in resource-limited settings.
- The Anemia Mukt Bharat program recommends POCT devices for anemia diagnosis.
Purpose of the Study:
- To evaluate the diagnostic accuracy of invasive and non-invasive digital hemoglobinometers (DHs) and Sahli's hemoglobinometer for anemia detection in children.
- To compare the performance of these POCT devices against a standard hematology analyzer.
Main Methods:
- A study involving 120 children aged 6-59 months was conducted in Haryana, India.
- Hemoglobin levels were measured using Sahli's hemoglobinometer, invasive DH, and non-invasive DH with capillary blood.
- Results were compared with hemoglobin levels obtained from a hematology analyzer using venous blood.
Main Results:
- Invasive DH (92.2%) and Sahli's hemoglobinometer (95.5%) demonstrated high sensitivity in diagnosing anemia.
- Non-invasive DH showed high specificity (96.7%) but low sensitivity (24.4%).
- Invasive DH provided the quickest results (2-3 minutes).
Conclusions:
- Invasive DHs are feasible and effective POCT devices for anemia diagnosis in young children.
- Sahli's hemoglobinometer is a viable option when trained personnel are available.
- Further research is needed to enhance the accuracy of non-invasive DH technology.
Introduction:
Globally around 47.4% of children and in India, 58% of children aged 6-59 months are anemic. Diagnosis of anemia in children using accurate technologies and providing adequate treatment is essential to reduce the burden of anemia. Point-of-care testing (POCT) devices is a potential option for estimation of hemoglobin in peripheral and field settings were the hematology analyzer and laboratory services are not available.
Objectives:
To access the validity of the POCTs (invasive and non-invasive devices) for estimation of hemoglobin among children aged 6-59 months compared with hematology analyzer.
Methods:
The study participants were enrolled from the pediatric outpatient department in Haryana, India, from November 2019 to January 2020. Hemoglobin levels of the study participants were estimated in Sahli's hemoglobinometer and invasive digital hemoglobinometers (DHs) using capillary blood samples. Hemoglobin levels in non-invasive DH were assessed from the finger/toe of the children. Hemoglobin levels measured in POCTs were compared against the venous blood hemoglobin estimated in the hematology analyzer.
Results:
A total of 120 children were enrolled. The mean (SD) of hemoglobin (g/dl) estimated in auto-analyzer was 9.4 (1.8), Sahli's hemoglobinometer was 9.2 (1.9), invasive DH was 9.7 (1.9), and non-invasive DH was 11.9 (1.5). Sahli's hemoglobinometer (95.5%) and invasive DH (92.2%) had high sensitivity for the diagnosis of anemia compared with non-invasive DH (24.4%). In contrast, non-invasive DH had higher specificity (96.7%) compared with invasive DH (83.3%) and Sahli's hemoglobinometer (70%). Invasive DH took the least time (2-3 min) for estimation of hemoglobin per participant, followed by Sahli's (4-5 min) and non-invasive DH (5-7 min).
Conclusion:
All three POCT devices used in this study are reasonable and feasible for estimating hemoglobin in under-5 children. Invasive DHs are potential POCT devices for diagnosis of anemia among under-5 children, while Sahli's can be considered as a possible option, where trained and skilled technicians are available. Further research and development are required in non-invasive DH to improve accuracy. Lay summaryIn India, anemia is a serious public health problem, where 58% of the children aged 6-59 months are anemic. Point-of-care testing (POCT) using digital hemoglobinometers (DHs) has been recommended as one of the key interventions by the Anemia Mukt Bharat program since 2018 in India. These POCT devices are easy to use, less invasive, can be carried to field, require minimal training and results are available immediately. Therefore this study assessed the validity of POCT devices-invasive DH, non-invasive DH and Sahli's hemoglobinometer among 6-59 months children in facility setting compared with the gold standard hematology analyzer. A total of 120 children under 6-59 months of age were enrolled from the pediatric outpatient department in Haryana, India, from November 2019 to January 2020. The (mean hemoglobin in g/dl) invasive (9.7) and non-invasive DH (11.9) overestimated hemoglobin value, while Sahli's (9.2) underestimated hemoglobin compared with hematology analyzer (9.4). Invasive DH (92.2%) and Sahli's hemoglobinometer (95.5%) reported high ability to correctly identify those with anemia compared with non-invasive DH (24.4%). In contrast, non-invasive DH (96.73%) had higher ability to correctly identify those without the anemia compared with invasive DH (83.3%) and Sahli's (70%).
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Errors occurring during blood pressure monitoring
Several factors...
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.