Related Experiment Video
Updated: Nov 24, 2025

A Simple Method for Automated Solid Phase Extraction of Water Samples for Immunological Analysis of Small Pollutants
Published on: January 1, 2016
Novel reversed-phase HPLC method for simultaneous determination of ethynodiol diacetate (EDA)/ethinyl estradiol (EE)
Srinivasa Reddy Chinta1, Vijetha Ponnam2, Naresh Kumar Katari3
1Department of Quality Operations, Novast Laboratories Ltd, Nantong, Jiangsu Province, China.
Abstract:
Ethynodiol diacetate (EDA) and ethinyl estradiol (EE) tablets are indicated to prevent pregnancy in women who use oral contraceptives as a contraception method. EDA and EE were separated by reversed-phase HPLC using Agilent ZORBAX SB-Phenyl column, 4.6 mm × 15 cm, 5 μm, using a gradient mixture of acetonitrile and Milli-Q water as mobile phase. The linearity and recovery were found in the range of 0.025-0.25 mg/mL and 0.05-0.18 mg/mL for EDA and 0.001-0.01 mg/mL and 0.002-0.007 mg/mL for EE, respectively. The method is validated according to the regulatory guidelines concerning system suitability, specificity, repeatability, recovery, linearity, robustness, and stability of the sample solution.
More Related Videos
13:35A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019