A comparative study between the two patch test systems Finn chambers and Finn chambers AQUA
Henrik Luu1, Martin Mowitz1, Magnus Bruze1
1Department of Occupational and Environmental Dermatology, Lund University, Skåne University Hospital, Malmö, Sweden.
Contact Dermatitis
|December 28, 2020
Summary
Finn Chambers AQUA (FCA) and Finn Chambers (FC) showed similar detection of positive contact allergies. Further research is needed to optimize FCA for reducing doubtful and irritant reactions in patch testing.
Area of Science:
- Dermatology
- Allergology
- Contact Dermatitis Research
Background:
- Finn Chambers AQUA (FCA) is an improved Finn Chambers (FC) test system with pre-fixed filter papers on a moisture-resistant adhesive patch.
- FCA simplifies the patch testing procedure by eliminating the need for separate filter papers and extra taping.
- This simplification may offer benefits to both patients and healthcare professionals performing patch tests.
Purpose of the Study:
- To compare the diagnostic accuracy of FCA and FC in detecting contact allergies.
- To evaluate potential differences in reaction types (positive, doubtful, irritant) between the two test systems.
Main Methods:
- A simultaneous patch testing study was conducted on 434 dermatitis patients.
- Ten common allergens were applied to patients using both FC and FCA test systems concurrently.
- Results from both systems were analyzed for differences in reaction detection.
Main Results:
- No significant differences were observed in the detection of positive contact allergy reactions between FC and FCA.
- FCA showed significantly more doubtful reactions to methylisothiazolinone, fragrance mix I, and hydroperoxides of linalool.
- FC showed significantly more doubtful reactions to nickel(II) sulfate, and FCA showed significantly more irritant reactions to formaldehyde.
Conclusions:
- The FC and FCA test systems demonstrate good agreement in identifying positive allergic contact dermatitis reactions.
- The observed differences in doubtful and irritant reactions suggest a need for further investigation into optimizing allergen dosage in FCA.
- Further research is warranted to refine the FCA system for improved clinical utility in allergy detection.