Related Experiment Video
Updated: Nov 24, 2025

Ye's Swing Technique for Small-incision Lenticule Extraction Surgery
Published on: June 27, 2025
[Results of corneal lenticule extraction for correction]
A V Doga1, S V Kostenev1, I A Mushkova1
1S.N. Fyodorov National Medical Research Center «MNTK «Eye Microsurgery», Moscow, Russia.
Abstract:
One of the main methods for correcting moderate and high myopia is corneal lenticule extraction. Ziemer is adding a new application to the FEMTO LDV Z8 platform - Lenticule CLEAR, designed to treat myopia by corneal lenticule extraction.
Abstract:
Purpose - to study the safety and effectiveness of corneal lenticule extraction using the Ziemer LDV Z8 device for the correction of moderate and high myopia.
Material And Methods:
The prospective study included 24 patients aged 26.8±4.6 years (24 eyes) who underwent surgeries for moderate and high myopia on the Ziemer LDV Z8 device at the S.N. Fyodorov National Medical Research Center «MNTK «Eye Microsurgery» in January 2020.
Results:
After correction of moderate and high myopia using the Ziemer LDV Z8 device, patients showed stable improvement in visual acuity one week after surgery, which remained throughout the postoperative period. Dynamic 1-month follow-up of patients operated on the Ziemer LDV Z8 platform for moderate and high myopia showed that the median value of uncorrected visual acuity one week after surgery was 0.9 [0.7; 1.0], and after one month - 1.0 [1.0; 1.0]. Patients operated on the Ziemer LDV Z8 platform had no intra- or postoperative complications, 79.2% of patients also had no loss of maximum corrected visual acuity. The mean corneal thickness before surgery was 555.9±28.2 μm, it significantly decreased to an average value of 464.8±26.9 μm one week after surgery (p<0.05). The cost of tissue was 17 microns per 1 diopter.
Conclusion:
The obtained results indicate the high effectiveness of the correction of moderate and high myopia using the Ziemer LDV Z8 device.
More Related Videos
05:46Author Spotlight: Advancements in Refractive Surgical Correction for Presbyopia and Exploring Postoperative Visual Acuity
Published on: September 20, 2024
05:14Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Related Concept Videos
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...