Related Experiment Video
Updated: Nov 23, 2025

A Method for Quantifying Upper Limb Performance in Daily Life Using Accelerometers
Published on: April 21, 2017
Validity of Tools to Measure Physical Activity in Older Adults Following Total Knee Arthroplasty
Abstract:
Few validated tools exist for measuring physical activity following total knee arthroplasty (TKA) despite the importance of returning to sufficient levels of physical activity post-TKA to achieve health benefits. This study examined the validity of two clinical measures-the Fitbit, a commercially available personal activity monitor, and the Community Healthy Activities Model Program for Seniors (CHAMPS), a self-report questionnaire-compared with a reference standard accelerometer, the SenseWearTM Armband (SWA). At 6-month post-TKA, 47 participants wore the Fitbit and SWA for 4 days and then completed the CHAMPS. Moderate-to-good correlation was observed between the Fitbit and SWA for steps (intraclass correlation coefficient [ICC] = .79), energy expenditure (ICC = .78), and energy expenditure <3 METS (ICC = .79). Poor-to-moderate correlation was observed between the CHAMPS and SWA (ICC = .43) with the questionnaire reporting lower daily energy expenditures than the SWA. Results showed that Fitbit may be a reasonable measurement tool to measure steps and energy expenditures in older adults following TKA.
More Related Videos
07:22Glycemic Impact on Knee Osteoarthritis Symptoms on Physical, Radiographic, and Inflammatory Markers among Individuals Aged 50 and Over with Diabetes
Published on: March 7, 2025
09:10The Lower Body Positive Pressure Treadmill for Knee Osteoarthritis Rehabilitation
Published on: July 22, 2019