Related Experiment Video
Updated: Nov 23, 2025

Frequency Mixing Magnetic Detection Scanner for Imaging Magnetic Particles in Planar Samples
Published on: June 9, 2016
Improved accuracy in high-frequency AC transport measurements in pulsed high magnetic fields
Hiroyuki Mitamura1, Ryuta Watanuki2, Erik Kampert3
1Institute for Solid State Physics, The University of Tokyo, Kashiwa 277-8581, Japan.
Abstract:
We show theoretically and experimentally that accurate transport measurements are possible even within the short time provided by pulsed magnetic fields. For this purpose, a new method has been devised, which removes the noise component of a specific frequency from the signal by taking a linear combination of the results of numerical phase detection using multiple integer periods. We also established a method to unambiguously determine the phase rotation angle in AC transport measurements using a frequency range of tens of kilohertz. We revealed that the dominant noise in low-frequency transport measurements in pulsed magnetic fields is the electromagnetic induction caused by mechanical vibrations of wire loops in inhomogeneous magnetic fields. These results strongly suggest that accurate transport measurements in short-pulsed magnets are possible when mechanical vibrations are well suppressed.
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences