Comparison study of the axial length measured using the new swept-source optical coherence tomography ANTERION and
Kook Young Kim1, Gon Soo Choi1, Min Seok Kang1
1Kim's Eye Hospital, Seoul, Korea.
Purpose:
To compare a biometer using swept-source optical coherence tomography (SS-OCT) with a partial coherence interferometry (PCI)-based biometer in measurements of two ocular biometry parameters, i.e., the axial length and anterior cornea curvature.
Methods:
We compared the two biometers SS-OCT (ANTERION, Heidelberg Engineering Inc., Heidelberg, Germany) and PCI (IOL Master, Carl Zeiss Meditec, Jena, Germany) in terms of the axial length (AL) and corneal curvature (K) measurements of 175 eyes. Paired t-tests were used to compare the two biometers. Agreement between the biometers was evaluated using the Bland-Altman method.
Results:
The mean age was 36.0 ± 25.6 years (range: 5 to 85 years). The mean axial length was 24.42 ± 0.13 mm for SS-OCT and 24.45 ± 0.14 mm for PCI. The mean corneal curvature was significantly different between the two biometry in flat K (K1) but not in steep K (K2). The limit of agreement was -0.15 to 0.21 in the axial length, -1.18 to 0.83 in K1, and -1.06 to 0.95 in K2. All above ocular biometric measurements between SS-OCT and PCI correlated significantly (Pearson's correlation, p<0.001).
Conclusions:
The axial length measured using SS-OCT is useful in clinical practice. It shows a good correlation and agreement with that measured using PCI. However, the axial length and corneal curvature measured using SS-OCT cannot be used interchangeably with that measured using PCI in clinical practice.