Related Experiment Video
Updated: Nov 23, 2025

Guided Endodontics: Three-Dimensional Planning and Template-Aided Preparation of Endodontic Access Cavities
Published on: May 24, 2022
Accuracy of root-end resection using a digital guide in endodontic surgery: An in vitro study
Li Peng1, Jing Zhao2, Zu-Hua Wang3
1Department of General Dentistry Ⅱ, Peking University School and Hospital of Stomatology, Beijing, China.
Background/Purpose:
It is difficult to achieve accurate root-end resection clinically. This in vitro study was conducted to evaluate the operation accuracy of a digital endodontic surgical guide.
Materials And Methods:
56 extracted maxillary anterior teeth were prepared for endodontic surgical models. The models were randomly divided into 4 groups equally according to the guide (with guide/no guide) and the operator (experienced/inexperienced). Endodontic microsurgeries were performed on models in each group. The deviations in length and angle of the root-end resection were measured based on the optical scanning data of the pre- and postoperative teeth. The general linear model was performed to analyze the effect of a guide on root-end resection deviation.
Results:
Using a guide, the mean length deviation for experienced/inexperienced operators reduced from 0.99 mm (95% CI [confidence interval, CI], 0.66-1.33 mm)/1.18 mm (95% CI, 0.50-1.86 mm) to 0.31 mm (95% CI, 0.20-0.42 mm)/0.31 mm (95% CI, 0.24-0.37 mm). The mean angle deviation for experienced/inexperienced operators reduced from 16.74° (95% CI, 10.61-22.86°)/15.06° (95% CI, 9.19-20.94°) to 5.04° (95% CI, 3.31-6.77°)/6.79° (95% CI, 4.91-8.67°). The difference was significant between procedures performed with and without a guide (P < 0.01).
Conclusion:
Application of the digital guide in vitro endodontic surgery could improve the accuracy of root-end resection.
More Related Videos
07:03Dynamic Navigation in Endodontics: Guided Access Cavity Preparation by Means of a Miniaturized Navigation System
Published on: May 5, 2022