Related Experiment Video
Updated: Nov 23, 2025

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Characterization of ferrous-agarose-xylenol gel dosimeter at 60Co γ-rays beam therapy unit
1Secondary Standard Dosimetry Laboratory, Nuclear Research Center of Algiers, 02 Boulevard Frantz Fanon, B.P.399, Algiers, Algeria. o.moussous@crna.dz.
Abstract:
Fricke gel dosimetry is a valuable technique used for recording 3D dose distribution in radiotherapy. Herein, we present the dosimetric characteristics of a synthesized ferrous-agarose-xylenol orange gel dosimeter in a clinical 60Co beam. Experimental data were obtained using a secondary standards dosimetry laboratory 60Co therapy unit. The dosimeter was calibrated using the ionization chamber as a reference, and its total mass attenuation coefficient, absorption spectrum, optical density-dose relationship, sensitivity, and dose rate dependency were evaluated. The potentiality of the ferrous-agarose-xylenol gel dosimeter was investigated to measure output factors for different field sizes. The gel dosimeter readings were measured using a spectrophotometer. The ferrous-agarose-xylenol gel dosimeter exhibited a linearity in the range of 3-15 Gy, indicating that the dosimeter is tissue-equivalent and dose rate-independent and yield reproducible results. The measured output factors and those published in the literature showed good agreement.
Related Concept Videos
Absorption of Radiation
Biological Effects of Radiation