Inter-vendor comparison of left atrial strain using layer specific strain analysis.
Aaisha Ferkh1,2,3, Luke Stefani2,3, Siddharth J Trivedi1,2,3
1Westmead Clinical School, University of Sydney, Sydney, NSW, Australia.
Vendor differences in left atrial strain (LAS) measurements using transthoracic echocardiogram (TTE) impact clinical use. GE mid-myocardial LAS underestimates strain compared to TomTec, while GE endocardial LAS shows no bias but greater variability. Serial LAS assessments require consistent vendor software.
Area of Science:
- Cardiology
- Medical Imaging
- Echocardiography
Background:
- Left atrial strain (LAS) assessed via transthoracic echocardiogram (TTE) is crucial for cardiovascular disease evaluation.
- Variability in LAS measurements between different software vendors hinders widespread clinical adoption.
Purpose of the Study:
- To compare LAS measurements between General Electric (GE) Echopac (mid-myocardial and endocardial) and TomTec software.
- To evaluate the impact of vendor-specific algorithms on LAS assessment in healthy individuals and paroxysmal atrial fibrillation (PAF) patients.
Main Methods:
- LAS (reservoir, contractile, conduit strain) was measured using GE Echopac (mid-myocardial and endocardial) and TomTec software in 88 healthy subjects and 40 PAF patients.
- Bland-Altman analysis was employed to assess agreement and systematic differences between vendor measurements.
Main Results:
- Both GE mid-myocardial and endocardial LAS correlated well with TomTec.
- GE mid-myocardial LAS systematically underestimated strain compared to TomTec.
- GE endocardial LAS showed no systematic bias but wider limits of agreement; reservoir strain demonstrated the best reproducibility.
Conclusions:
- Routine GE Echopac mid-myocardial LAS underestimates values obtained with TomTec.
- GE endocardial LAS mitigates bias but increases measurement variability.
- Consistent use of the same vendor system is recommended for serial LAS measurements to ensure reliability.