CT artifacts after contrast media injection in chest imaging: evaluation of post-processing algorithms, virtual
Amit Gupta1,2, Verena Carola Obmann1,2,3, Michelle Jordan1,2
1Department of Radiology, University Hospitals Cleveland Medical Center, Cleveland, OH, USA.
Background:
After injection into a brachial vein, high contrast media concentration in axillary and subclavian veins can cause artifacts that impair diagnostic utility. This study assessed artifact reduction by artifact-reduction-algorithms (ARA) and virtual-monoenergetic-images (VMI), as well as their combination (VMIARA) compared to conventional CT-images (CI).
Methods:
Forty-six spectral-detector-CT (SDCT) examinations of patients that received ARA-reconstructions due to perivenous-artifacts were included in this retrospective study. CI, ARA, VMI, and VMIARA (range: 70-200 keV) were reconstructed. Objective analysis was performed with ROI-based assessment of mean and standard deviation of attenuation (HU) in hypo- and hyperdense artifacts and impaired muscle and arteries as well as artifact-free reference-tissue. Extent of artifact reduction, assessment of surrounding soft tissue and vessels, and appearance of new artifacts were rated visually by two radiologists.
Results:
Hypo- and hyperdense artifacts showed significant improvement as evidenced by decreasing attenuation differences between artifact impaired and artifact-free reference tissue in ARA, VMI ≥80 keV, and VMIARA between 70-200 keV (e.g., CI/ARA/VMI100keV/VMIARA100keV: hypodense artifacts, (-)264.8±150.9/(-)87.1±78.9/(-)48.6±64.6/9.9±63.9 HU; P<0.001); hyperdense artifacts, 164.2±51.1/82.1±73.2/7.9±34.7/(-)17.3±50.7 HU; P<0.001). Artifacts impairing surrounding muscle and arteries were also reduced by all three approaches. In visual assessment, ARA, VMI ≥100 keV, and VMIARA between 70-200 keV also showed significant artifact reduction and improved assessment; however, for assessment of arteries improvement was not significant using ARA alone. New artifacts were reported, particularly at higher keV-values.
Conclusions:
In presence of perivenous-artifacts, ARA, VMI and their combination allow for significant artifact reduction; however, their combination and VMI as a standalone approach yielded best results and should therefore be used, if available.
More Related Videos
05:32Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
12:15Tissue Preparation Techniques for Contrast-Enhanced Micro Computed Tomography Imaging of Large Mammalian Cardiac Models with Chronic Disease
Published on: February 8, 2022
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...