Impact of preset and postset adaptive statistical iterative reconstruction-V on image quality in nonenhanced abdominal-pelvic CT on wide-detector revolution CT
Zheng Zhu1, Yanfeng Zhao1, Xinming Zhao1
1Department of Diagnostic Radiology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
Background:
Adaptive statistical iterative reconstruction-V technique (ASIR-V) is usually set at different strengths according to the different clinical requirements and scenarios encountered when setting scanning protocols, such as setting a more aggressive tube current reduction (defined as preset ASIR-V). Reconstruction with ASIR-V is useful after scanning using image algorithms to improve image quality (defined as postset ASIR-V). The aim of this study was to investigate the quality of images reconstructed with preset and postset ASIR-V, using the same noncontrast abdominal-pelvic computed tomography (CT) protocols in the same individual on a wide detector CT.
Methods:
We prospectively enrolled 141 patients. The scan protocols in Groups A-E were 0%, 20%, 40%, 60%, and 80% preset ASIR-V, respectively, in the 256 wide-detector row Revolution CT (GE Healthcare, Waukesha, WI, USA). Each group was further divided into 5 subgroups with 0%, 20%, 40%, 60%, and 80% postset ASIR-V, respectively. The 64-detector Discovery 750 HDCT (GE, USA) was used for Group F as a control group, using 0%, 20%, 40%, 60%, and 80% ASIR, respectively. Image noise was measured in the spleen, aorta, and muscle. The CT attenuation and image noise were analyzed using the paired t-test; analysis of variance and post hoc multiple comparisons were made using the Student-Newman-Keuls (SNK) method.
Results:
The CT attenuation in Groups A-F exhibited no significant difference between subgroups in three organs (P>0.05). Only with increasing preset ASIR-V% (Groups A to E), did the image noise decrease, except in Group B in the aorta and muscle (NoiseB > NoiseA, PmuscleA&B=0.233, PaortaA&B=0.796). Only with increasing postset ASIR-V or ASIR% (Groups A and F), did the image noise decrease in the three organs. After preset and postset ASIR-V were combined, with preset ASIR-V% being equal to postset ASIR-V%, the image become similar to the corresponding preset ASIR-V part with the line of postset ASIR-V 0% (baseline of each group). When preset ASIR-V% was greater than the postset ASIR-V%, the image noise was higher than the baseline of each group. When preset ASIR-V% was less than the postset ASIR-V%, the image noise was lower than the baseline of each group. The radiation dose from B to E decreased from 11.2% to 57.1%. The CT dose index volume (CTDIvol) and dose length product (DLP) in Group F were significantly higher than those in Group A.
Conclusions:
Using both preset and postset ASIR-V allows dose reduction, with a potential to improve image quality only when postset ASIR-V% is higher than or equal to preset ASIR-V%. The image quality depends on postset ASIR-V%, whereas the decrease of radiation dose depends on preset ASIR-V%.
More Related Videos
05:32Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...