Related Experiment Video
Updated: Nov 22, 2025

A Millimeter Scale Flexural Testing System for Measuring the Mechanical Properties of Marine Sponge Spicules
Published on: October 11, 2017
Pressure unit inconsistency in the Medical Measurement Systems ASCII-file
1College of Medicine & Public Health, Flinders University, Adelaide, SA, Australia.
Background:
To apply novel analysis methods for processing high-resolution pressure-impedance motility parameters, raw data acquired by commercial motility systems are exported and analyzed externally using purposed-designed software.
Methods:
Data acquired by the Medical Measurement Systems (MMS) motility equipment are exported from MMS in different file formats.
Key Results:
MMS studies that are exported to ASCII-file format will be in cmH2O units even though acquired and displayed by the MMS system in mmHg units (1 mmHg would be exported as 1.36 cmH2O).
Conclusions & Inferences:
Due to this issue, investigators may unknowingly generate pressure data in incorrect units, data exported from MMS in ASCII-file format need to be divided by a factor 1.36 to convert cmH2O to mmHg.
Related Concept Videos
Measurement: Standard Units
Units of Measurement
There are various well-known historical measurement systems, such as the Babylonian system, the Roman system, the Egyptian system, the Olympian system, the British system, and the Indus...
SI Units: 2019 Redefinition
A standard set of units has been defined...
Conversion of Units
The first step in the unit conversion is to list the given units and the units required...
Units and Standards of Measurement
Measurements of physical quantities are...
International System of Units
Prefixes are used to define both larger and smaller quantities in the SI system. For example, milli, micro, and nano define smaller quantities, while kilo, mega, and giga are used to define...