Comparison between two multimodal imaging platforms: Nidek Mirante and Heidelberg Spectralis
Kimberly Spooner1, Long Phan1,2, Mariano Cozzi3
1Sydney Retina, Level 13, Park House, 187 Macquarie Street, Sydney, NSW, 2000, Australia.
Purpose:
To investigate the reliability and comparability of retinal measurements obtained with spectral-domain optical coherence tomography (OCT), optical coherence tomography angiography (OCTA), confocal scanning laser ophthalmoscopy (cSLO) colour images, and fundus autofluorescence (FAF) between two multimodal imaging platforms in eyes with macular pathology and normal, healthy volunteers.
Methods:
This cross-sectional, multi-centre, instrument validation study recruited 94 consecutive subjects. All participants underwent a dilated examination and were scanned consecutively on the Heidelberg Spectralis (Heidelberg Engineering, Heidelberg, Germany) and Nidek Mirante (Nidek Co. Ltd., Gamagori, Japan) devices. Agreement between device images were evaluated from measures of the central retinal thickness (CRT), presence of segmentation and fixation imaging artefacts (IA), foveal avascular zone (FAZ) measurements; as well as sensitivity and specificity values from the detection of atrophy on fundus autofluorescence (FAF), drusen, subretinal drusenoid deposits, geographic atrophy, epiretinal membrane, fibrosis and haemorrhage on multicolour imaging, and agreement between devices and groups.
Results:
Compared with reference clinical examination, sensitivity values for the identification of retinal features using sole device images ranged from 100% for epiretinal membranes to 66.7% for subretinal drusenoid deposits (SSD). Mean absolute difference for CRT between OCT devices was 3.78 μm (95% confidence interval [CI]: - 21.39 to 28.95, P = 0.809). Differences in the superficial and deep capillary plexus FAZ area on OCTA between devices were not statistically significant (P = 0.881 and P = 0.595, respectively). IAs were significantly increased in the presence of macular pathology.
Conclusion:
Comparison of retinal measurements between the OCT devices did not differ significantly. Common ultrastructural biomarkers of multiple macular pathologies were identified with high sensitivities and specificities, with good agreement between graders, indicating that they can be identified with comparable confidence in retinal imaging between the two devices.
Related Concept Videos
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...