Iodine determination in animal feed by inductively coupled plasma mass spectrometry - results of a collaborative

Ana Jerše1, Heidi Amlund1, Rie R Rasmussen1

  • 1National Food Institute, Technical University of Denmark , Kgs. Lyngby, Denmark.

A collaborative study was conducted in order to fully validate the performance characteristics and to evaluate the suitability of a method for determination of iodine in animal feed. The method consists of an alkaline extraction in tetramethylammonium hydroxide (TMAH) solution followed by the determination of iodine by inductively coupled plasma-mass spectrometry (ICP-MS). The method was validated for different types of feed and feed materials with a broad concentration range of 0.65-622 mg I/kg. Good agreement was found between the overall mean mass fraction values from the collaborative trial (13.8 ± 1.3 mg I/kg and 0.657 ± 0.228 mg I/kg) and the values previously determined in proficiency tests for two of the test materials (12.65 ± 2.47 mg I/kg and 0.72 ± 0.22 mg I/kg) indicating satisfactory accuracy of the method. Reproducibility standard deviations were between 7.85% and 34.65% and the HorRat values were under the acceptable limit of 2 so the between-laboratory precision was considered acceptable. Based on the statistical evaluation of the results it was concluded that the method is suitable for its intended purpose; it has been accepted as European Standard EN17050:2017 by the European Committee for Standardisation (CEN).

Related Concept Videos

Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview01:19

Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview

In inductively coupled plasma–mass spectrometry (ICP–MS), an inductively coupled plasma (ICP) torch is used as an atomizer and ionizer. Solid samples are dissolved and volatilized before being introduced into the high-temperature argon plasma, while solution samples are nebulized and passed through the high-temperature argon plasma. Plasma dissociates the analytes and ionizes their component atoms to form a mixture of positive ions and molecular species. The positive ions are then...
1.3K
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle01:19

Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle

Inductively coupled plasma (ICP) is the most widely used plasma source in atomic emission spectroscopy (AES), also known as Inductively Coupled Plasma Optical Emission Spectroscopy (ICP-OES). The ICP source, or torch, consists of three concentric quartz tubes with argon gas flowing through them. A spark from a Tesla coil initiates the ionization of argon, generating a high-temperature plasma.
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
1.3K
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences01:20

Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences

Inductively coupled plasma–mass spectrometry (ICP–MS) is a highly selective and sensitive technique for accurate elemental analysis. Though the analysis of ICP–MS mass spectra is comparatively straightforward, it is affected by spectroscopic and non-spectroscopic interferences. Spectroscopic interferences arise when the plasma contains ionic species with an m/z value the same as the analyte ion. Spectroscopic interference can be categorized as isobaric, polyatomic ions, and...
892
Redox Titration: Iodimetry and Iodometry01:23

Redox Titration: Iodimetry and Iodometry

Iodometry and iodimetry are analytical methods used to determine the concentration of oxidizing or reducing agents using iodine. In iodometric titrations, the oxidizing analyte solution is usually acidified and treated with an excess of iodide ions, which generates an equivalent amount of iodine in equilibrium with triiodide. The released iodine is subsequently titrated directly against a standardized reducing agent. As the dilute iodine color becomes pale yellow, a few drops of freshly...
4.1K
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation01:26

Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation

Inductively coupled plasma (ICP) is the common plasma source used in atomic emission spectroscopy (AES), a technique that detects and analyzes various elements in a sample. This method is often called inductively coupled plasma atomic emission spectroscopy (ICP-AES).
There are three main types of inductively coupled plasma atomic emission spectroscopy  (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
465