Cone-beam CT versus Multidetector CT in Postoperative Cochlear Implant Imaging: Evaluation of Image Quality and
R A Helal1, R Jacob2, M A Elshinnawy3
1From the Radiodiagnosis Department (R.A.H., M.A.E., A.I.O., T.T.A.), Ain Shams University, Cairo, Egypt raniahelal@med.asu.edu.eg.
Background And Purpose:
Cone-beam CT is being increasingly used in head and neck imaging. We compared cone-beam CT with multidetector CT to assess postoperative implant placement and delineate finer anatomic structures, image quality, and radiation dose used.
Materials And Methods:
This retrospective multicenter study included 51 patients with cochlear implants and postoperative imaging via temporal bone cone-beam CT (n = 32 ears) or multidetector CT (n = 19 ears) between 2012 and 2017. We evaluated the visualization quality of single electrode contacts, the scalar position of the electrodes, cochlear walls, mastoid facial canal, metallic artifacts (using a 4-level visual score), and the ability to measure the insertion angle of the electrodes. The signal-to-noise ratio and radiation dose were also evaluated.
Results:
Cone-beam CT was more sensitive for visualizing the scalar position of the electrodes (P = .046), cochlear outer wall (P = .001), single electrode contacts (P < .001), and osseous spiral lamina (P = .004) and had fewer metallic artifacts (P < .001). However, there were no significant differences between both methods in visualization of the modiolus (P = .37), cochlear inner wall (P > .99), and mastoid facial canal wall (P = .07) and the ability to measure the insertion angle of the electrodes (P > .99). The conebeam CT group had significantly lower dose-length product (P < .001), but multidetector CT showed a higher signal-to-noise ratio in both bone and air (P = .22 and P = .001).
Conclusions:
Cone-beam CT in patients with cochlear implants provides images with higher spatial resolution and fewer metallic artifacts than multidetector CT at a relatively lower radiation dose.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
05:49Author Spotlight: Advancing CBCT and Digital Dental Image Integration with AI-Assisted Digitization
Published on: February 23, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT