Simultaneous Determination of Nitrite and Azide Ions in Valsartan
1Department of Pharmacy, Children's Hospital of Nanjing Medical University, Nanjing 210008, China.
Objective:
This study aimed to establish an ion chromatography for the simultaneous determination of nitrite and azide ions in valsartan.
Methods:
Dionex IonpacAG18 (4 × 50 mm2) and Dionex IonpacAS18 (4 × 250 mm2) were used as the 2D guard column and analytical column, respectively. Dionex AMGTMC18 (4 × 30 mm2) was used as a 1D pre-separation column and Dionex UTAC-ULP1 as the concentration column. KOH solution generated by an online eluent generator (EG) was used as the 2D mobile phase, and gradient elution was performed at 1 mL/min. Water and acetonitrile were used as 1D mobile phase, and gradient elution was performed at 0.5 mL/min. Suppressed conductivity detector (suppressor, ASRS 300 4 mm) was used for detection. The column temperature was 30°C, while that of the detector was 35°C.
Results:
This method showed satisfactory reproducibility and recovery. The linear range and the correlation coefficient for both nitrite and azide ions were 0.01-0.2 mg/L and 0.9998, respectively. However, the recovery rate was 85-115% for nitrite and 90-110% for azide ions.
Conclusion:
This method was suitable for the simultaneous determination of nitrite and azide ions in valsartan and other sartan drugs.