Related Experiment Video
Updated: Nov 22, 2025

Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
Biomechanical Comparison between Volar Plate Fixator and Nonbridge External Wrist Fixator
M A Hajianpour1, T Sano2, E A Ouellette3
1Total Orthopaedic Care, Lauderdale Lakes, FL 33313.
The nonbridging fixator (NBX) offers comparable fracture stability to volar plates for complex distal radius fractures, with better alignment restoration and minimal impact on range of motion.
Area of Science:
- Orthopedic Surgery
- Biomechanical Engineering
- Trauma Care
Background:
- Intra-articular distal radius fractures present challenges for nonoperative management.
- Open Reduction and Internal Fixation (ORIF) involves implants that remain post-healing.
- External fixation offers stabilization without long-term implants.
Purpose of the Study:
- To compare the nonbridging fixator (NBX) with volar plate fixation (VPS) for 5-fragment, OTA 23 C3.2 distal radius fractures.
- To evaluate reduction quality, fixation rigidity, and range of motion (ROM).
Main Methods:
- A 5-part distal radius fracture model was created in 10 cadaver arms.
- Arms were randomly assigned to NBX or VPS fixation.
- Fluoroscopic imaging assessed ROM; axial loading determined failure strength.
Main Results:
- NBX demonstrated superior reduction in radial tilt (13.8° vs 6.3°) and volar tilt (26.3° vs 14.0°).
- NBX showed better restoration of radial length (3.4 mm vs 1.9 mm).
- NBX fixation had minimal effect on ROM, while VPS increased it post-fracture; NBX peak load was 925 N vs 2152 N for VPS.
Conclusions:
- The NBX provides adequate strength and superior alignment restoration compared to VPS for this fracture model.
- NBX offers a viable alternative to ORIF, minimizing long-term implant burden.
- NBX fixation preserves joint function with minimal ROM compromise.
More Related Videos
07:41An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
08:20A Reliable and Reproducible Critical-Sized Segmental Femoral Defect Model in Rats Stabilized with a Custom External Fixator
Published on: March 24, 2019