Error grid analysis for risk management in the difference between invasive and noninvasive blood pressure
Takashi Juri1, Koichi Suehiro2, Akihiro Uchimoto1
1Department of Anesthesiology, Osaka City University Graduate School of Medicine, 1-5-7 Asahimachi, Abenoku, Osaka City, Osaka, 545-8586, Japan.
Purpose:
Invasive arterial blood pressure (IAP) and noninvasive blood pressure (NIBP) measurements are both common methods. Recently, a new method of error grid analysis was proposed to compare blood pressure obtained using two measurement methods. This study aimed to compare IAP and NIBP measurements using the error grid analysis and investigate potential confounding factors affecting the discrepancies between IAP and NIBP.
Methods:
Adult patients who underwent general anesthesia in the supine position with both IAP and NIBP measurements were retrospectively investigated. The error grid analyses were performed to compare IAP and NIBP. In the error grid analysis, the clinical relevance of the discrepancies between IAP and NIBP was evaluated and classified into five zones from no risk (A) to dangerous risk (E).
Results:
Overall, data of 1934 IAP/NIBP measurement pairs from 100 patients were collected. The error grid analysis revealed that the proportions of zones A-E for systolic blood pressure were 96.4%, 3.5%, 0.05%, 0%, and 0%, respectively. In contrast, the proportions for mean blood pressure were 82.5%, 16.7%, 0.8%, 0%, and 0%, respectively. The multiple regression analysis revealed that continuous phenylephrine administration (p = 0.016) and age (p = 0.044) were the significant factors of an increased clinical risk of the differences in mean blood pressure.
Conclusions:
The error grid analysis indicated that the differences between IAP and NIBP for mean blood pressure were not clinically acceptable and had the risk of leading to unnecessary treatments. Continuous phenylephrine administration and age were the significant factors of an increased clinical risk of the discrepancies between IAP and NIBP.
More Related Videos
14:28Software for Analysis of Heart Rate and Blood Pressure Time-series Data from the Valsalva Maneuver
Published on: June 27, 2025
09:56Implantation of Combined Telemetric ECG and Blood Pressure Transmitters to Determine Spontaneous Baroreflex Sensitivity in Conscious Mice
Published on: February 14, 2021
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)