Quantification of reservoir bags as airway pressure-limiting devices in a nonrebreathing system
Tamas D Ambrisko1, Arnon Gal1, Jesus N Sarol2
1Department of Veterinary Clinical Medicine, College of Veterinary Medicine, University of Illinois at Urbana-Champaign, Urbana, IL, USA.
Objective:
To evaluate the influence of reservoir bag types, volumes and previous use on the peak pressures (Pmax) and the times to develop 30 cmH2O pressure (P30) within a nonrebreathing system with a closed adjustable pressure-limiting (APL) valve.
Study Design:
In vitro study using three-way factorial design with repeated measure on one factor.
Subjects:
A total of 75 new anesthesia reservoir bags (five types, three volumes, five bags from each type × volume). The bag types were reusable latex (RL), disposable latex (DL) and three disposable neoprene (DN-1, DN-2 and DN-3).
Methods:
Each bag was tested three times (treatments): new, after prestretching and 1 week later. The bags were attached to a Bain system and anesthesia machine with closed APL valve and patient port with O2 flow 2 L minute-1 until Pmax was reached. The Pmax and time to reach P30 values were determined from recorded pressure traces. General linear mixed model analysis was used to examine the effects of bag type, volume and treatment. One-sided 95% upper prediction limits of Pmax were calculated to test the null hypothesis that predicted Pmax of new bags would be ≥ 50 cmH2O for each factor combination.
Results:
RL bags were the least and DN-3 bags were the most compliant. Prestretching increased compliance in all bag types. Smaller bags of RL, DL and DN-1 were less compliant than larger ones. The predicted Pmax values were < 50 cmH2O only for DN-3 bags after prestretching. The time to reach P30 was critically low when using 0.5 L bags (median 17 seconds).
Conclusions And Clinical Relevance:
To minimize the risk of barotrauma, highly compliant reservoir bags (e.g. DN-3) are recommended and reusable bags should be avoided. Bags should be prestretched before first use, 0.5 L bags should be avoided and fresh gas flow minimized.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Respiratory Capacities
One key metric is the Inspiratory Capacity (IC), which represents the maximum amount of air that can be inhaled with full effort. IC is calculated by summing the tidal volume and inspiratory reserve volume, typically ranging from 2.4 to 3.6 liters.
The Functional Residual Capacity (FRC) represents the air in the...
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...