Cardiac Output Evaluation on Septic Shock Patients: Comparison between Calibrated and Uncalibrated Devices during

Paolo Persona1, Ilaria Valeri2, Elisabetta Saraceni3

  • 1UOC Institute of Anaesthesia and Intensive Care Unit, University Hospital of Padua, 35121 Padua, Italy.

Insights

The MostCare system shows good agreement with the PiCCO transpulmonary thermodilution method for measuring cardiac output in septic shock patients. This non-invasive device offers a reliable option for monitoring hemodynamic changes in emergency settings.

Area of Science:

  • Critical Care Medicine
  • Cardiovascular Physiology
  • Medical Devices

Background:

  • Accurate cardiac output (CO) monitoring is crucial in septic shock management.
  • Existing non-invasive methods for CO measurement in sepsis lack reliability.
  • The MostCare device analyzes arterial waveforms for beat-to-beat CO estimation.

Purpose of the Study:

  • To evaluate the agreement between the MostCare system and transpulmonary thermodilution (PiCCO) for CO measurement in septic shock patients.
  • To assess the ability of MostCare to track CO changes during norepinephrine reduction.
  • To determine the reliability of MostCare under varying hemodynamic conditions, including volume expansion.

Main Methods:

  • A prospective study involving 40 septic shock patients requiring norepinephrine.
  • Comparison of CO measurements between MostCare and PiCCO (calibrated transpulmonary thermodilution).
  • Data collection at two time points (T1 and T2) during norepinephrine reduction, with separate analysis for volume expansion events.
  • Statistical analysis included Bland-Altman for static agreement and polar plots for trending.

Main Results:

  • At T1, mean CO was 5.38 L/min with a percentage error (PE) of 22.6% between MostCare and PiCCO.
  • At T2, mean CO was 5.44 L/min with a PE of 27%.
  • When accounting for volume expansion, PE ranged from 17% to 30%.
  • Polar plot analysis supported the trending ability of the MostCare system.

Conclusions:

  • The MostCare system demonstrates good agreement with PiCCO for CO monitoring in septic patients.
  • MostCare reliably tracks hemodynamic variations during norepinephrine reduction and volume changes.
  • MostCare presents a rapid, reliable, and potentially valuable tool for hemodynamic monitoring in emergency settings.

Related Concept Videos

Equipments Used To Measure Blood Pressure01:30

Equipments Used To Measure Blood Pressure

Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
2.8K
Cardiac Output I:Effect of Heart Rate on Cardiac Output01:19

Cardiac Output I:Effect of Heart Rate on Cardiac Output

Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
2.0K
Assessing Blood pressure using a doppler ultrasound01:19

Assessing Blood pressure using a doppler ultrasound

To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
2.0K
Cardiac Output II: Effect of Stroke Volume on Cardiac Output01:22

Cardiac Output II: Effect of Stroke Volume on Cardiac Output

Cardiac output (CO), the amount of blood the heart pumps per minute, is a parameter in cardiovascular physiology determined by stroke volume and heart rate. Stroke volume, the amount of blood pushed from one of the ventricles per heartbeat, is influenced by preload, afterload, and contractility.
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
2.4K