Cardiac Output Evaluation on Septic Shock Patients: Comparison between Calibrated and Uncalibrated Devices during
Paolo Persona1, Ilaria Valeri2, Elisabetta Saraceni3
1UOC Institute of Anaesthesia and Intensive Care Unit, University Hospital of Padua, 35121 Padua, Italy.
Insights
The MostCare system shows good agreement with the PiCCO transpulmonary thermodilution method for measuring cardiac output in septic shock patients. This non-invasive device offers a reliable option for monitoring hemodynamic changes in emergency settings.
Area of Science:
- Critical Care Medicine
- Cardiovascular Physiology
- Medical Devices
Background:
- Accurate cardiac output (CO) monitoring is crucial in septic shock management.
- Existing non-invasive methods for CO measurement in sepsis lack reliability.
- The MostCare device analyzes arterial waveforms for beat-to-beat CO estimation.
Purpose of the Study:
- To evaluate the agreement between the MostCare system and transpulmonary thermodilution (PiCCO) for CO measurement in septic shock patients.
- To assess the ability of MostCare to track CO changes during norepinephrine reduction.
- To determine the reliability of MostCare under varying hemodynamic conditions, including volume expansion.
Main Methods:
- A prospective study involving 40 septic shock patients requiring norepinephrine.
- Comparison of CO measurements between MostCare and PiCCO (calibrated transpulmonary thermodilution).
- Data collection at two time points (T1 and T2) during norepinephrine reduction, with separate analysis for volume expansion events.
- Statistical analysis included Bland-Altman for static agreement and polar plots for trending.
Main Results:
- At T1, mean CO was 5.38 L/min with a percentage error (PE) of 22.6% between MostCare and PiCCO.
- At T2, mean CO was 5.44 L/min with a PE of 27%.
- When accounting for volume expansion, PE ranged from 17% to 30%.
- Polar plot analysis supported the trending ability of the MostCare system.
Conclusions:
- The MostCare system demonstrates good agreement with PiCCO for CO monitoring in septic patients.
- MostCare reliably tracks hemodynamic variations during norepinephrine reduction and volume changes.
- MostCare presents a rapid, reliable, and potentially valuable tool for hemodynamic monitoring in emergency settings.
Abstract:
There are no reliable, non-invasive methods to accurately measure cardiac output (CO) in septic patients. MostCare (Vytech Health™, Vygon, Padova, Italy), is a beat-to-beat, self calibrated method for CO measurement based on continuous analysis of reflected arterial pressure waveforms. We enrolled 40 patients that were suffering from septic shock and requiring norepinephrine infusion to target blood pressure in order to to evaluate the level of agreement between a calibrated transpulmonary thermodilution device (PiCCO System, Pulsion Medical Systems, Feldkirchen, Germany) and the MostCare system in detecting and tracking changes in CO measurements related to norepinephrine reduction in septic shock patients,. PiCCO was connected to a 5 Fr femoral artery catheter and to a central venous catheter. System calibration was performed with 15 mL of cold saline injection over about 3 s. The MostCare device was connected to the artery catheter to analyze the arterial waveform. Before reducing norepinephrine infusion, the PiCCO system was calibrated, the MostCare waveform was optimized, and the values of the complete hemodynamic profile were recorded (T1). Norepinephrine infusion was then reduced by 0.03 mcg/Kg/min. After 30 min, a new calibration of PiCCO system and a new record on both monitors were performed (T2). Static measurements agreements were assessed using the Bland-Altman test, while trending ability was investigated using polar plot analysis. If volume expansion occurred, then related data were separately analyzed. At T1 mean the CO was 5.38 (SD 0.60) L/min, the mean difference was 0.176 L/min, the limits of agreement (LoA) was +1.39 and -1.04 L/min, and the percentage error (PE) was 22.6%; at T2 the mean CO was 5.44 (SD 0.73) L/min, the mean difference was 0.053 L/min, the LoA was +1.51 and -1.40, and the PE was 27%. After considering the volume expansion between T1 and T2, the mean CO at T1 was 5.39 L/min (SD 0.47), the LoA was +1.09 and -0.78 L/min, and the percentage error (PE) was 17%; at T2 the mean CO was 5.35 L/min (SD 0.81), the LoA was +1.73 and -1.52 L/min, and the PE was 30%. The polar plot diagram seems to confirm the trending ability of MostCare system versus the reference method. In septic patients, when the arterial waveform is accurate, MostCare and PiCCO transpulmonary thermodilution exhibit good agreement even after the reduction of norepinephrine and changes in vascular tone or volume expansion. MostCare could be a rapid to set, reliable, and useful tool to monitor hemodynamic variations in septic patients in emergency contexts where thermodilution methods or other advanced systems are not easily available.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...