Measuring 3D In-vivo Shoulder Kinematics using Biplanar Videoradiography
Published on: March 12, 2021
Influence of Radiographic Viewing Perspective on Glenoid Inclination Measurement
Peter N Chalmers1, Thomas Suter2, Matthijs Jacxsens3
1Department of Orthopaedic Surgery, University of Utah, Salt Lake City, Utah.
Introduction:
The purposes of this study were to determine (1) whether glenoid inclination (GI) could be accurately measured on plain radiographs as compared to a gold-standard 3-dimensional (3D) measure and (2) whether GI could be reliably measured on plain radiographs.
Materials And Methods:
Digitally reconstructed radiographs (DRRs) were made from 3D computed tomography reconstructions of 68 normal cadaver scapulae. DRRs were made in a variety of viewing angles. Inclination was measured on these DRRs. These measurements were also made using a gold-standard 3D method. Measurements were made by 2 orthopedic surgeons and 1 surgeon twice, to calculate interrater and intrarater intraclass correlation coefficients (ICCs).
Results:
The gold-standard 3D β was 83 ± 5° (72°-98°). On neutral plain radiographs, the mean ± standard deviation 2D β angle was 80 ± 6° (range, 66°-99°). With regard to accuracy, the 2D β angle was significantly different from the 3D β angle, with the 2D β underestimating the 3D β by 5° (95% confidence intervals -1 to 12). With regard to reliability, interrater ICCs for 2D β with a neutral viewing angle was 0.79. Two-dimensional β varied widely with viewing angle from 0.24 to 0.88. Interrater ICCs for the 3D method was 0.83 (0.60-0.92). Intrarater ICCs for all 3 techniques were high (>0.91).
Conclusions:
Two-dimensional radiographic GI measurement is not accurate, as it underestimates the 3D value by an average of 5° when compared to the gold-standard 3D measurement. GI 2D measurement reliability varies with viewing angle on plain radiographs and thus to accurately and reliably measure inclination 3D imaging is necessary.