Cardiac 123I-MIBG planar heart to mediastinum ratios depend on patient size; phantom studies suggest SPECT-CT could
Gemma Roberts1,2, Jim J Lloyd1,2, George S Petrides2
1Institute of Neuroscience, Newcastle University, Biomedical Research Building, Campus for Ageing and Vitality, Westgate Road, Newcastle Upon Tyne. NE4 6BE, United Kingdom.
Introduction:
Planar 123I-MIBG (meta-iodobenzylguanidine) cardiac imaging is listed as an indicative biomarker in the 2017 international consensus criteria for the diagnosis of dementia with Lewy bodies. There has been very little research into the relationship between apparent cardiac uptake and patient size, or in the possible advantage of attenuation and scatter corrected SPECT-CT compared to planar imaging. We aimed to evaluate this in both a chest phantom and in older adults with normal cognition.
Materials And Methods:
An anthropomorphic chest phantom was filled with 123I solution using activities typical of healthy subjects. The phantom was scanned on a Siemens Intevo gamma camera with MELP collimators using both planar and SPECT-CT techniques. Further scans were acquired with a PMMA chest plate added, then water filled plastic breasts. The SPECT-CT images were reconstructed using a resolution recovery OSEM method with and without attenuation and scatter correction (ACSC) applied. Twenty-nine adults over 60 years of age (mean 75.2 ± 8.3 years) underwent planar cardiac MIBG imaging, followed by SPECT-CT. SPECT images were reconstructed as above. Heart-to-mediastinum ratios (HMRs) were calculated for planar and SPECT images.
Results:
Phantom planar HMR decreased by 20% with the PMMA chest plate added; 39% with plate and breasts. ACSC SPECT cardiac counts showed less dependence on phantom size than SPECT without ACSC (3% versus 37%). The body mass indices (BMI) of the older adults ranged from 22 to 38. There was a significant linear relationship between planar HMR and BMI (R2 = 0.44, p<0.01), but not for ACSC SPECT. However, there was no significant difference between the slopes for planar and ACSC SPECT (p = 0.11).
Conclusion:
Planar cardiac 123I-MIBG HMR results are correlated with BMI. Phantom results suggest that ACSC SPECT can correct for patient size. A large patient population or clinical database would be required to demonstrate a clinical effect.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System IV: CMRI
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...