Impact of a parallel magnetic field on radiation dose beneath thin copper and aluminum foils
Radim Barta1, Andrei Ghila1, Satyapal Rathee1,2
1Department of Oncology, Medical Physics Division, University of Alberta, 11560 University Avenue, Edmonton, Alberta T6G 1Z2, Canada.
Purpose:
The RF coils for magnetic resonance image guided radiotherapy (MRIgRT) may be constructed using thin and/or low-density conductors, along with thinner enclosure materials. This work measures the surface dose increases for lightweight conductors and enclosure materials in a magnetic field parallel to a 6 MV photon beam.
Methods:
Aluminum and copper foils (9-127 μm thick), as well as samples of polyimide (17 μm) and polyester (127 μm) films are positioned atop a polystyrene phantom. A parallel plate ion chamber embedded into the top of the phantom measures the surface dose in 6 MV photon beam. Measurements (% of dose at the depth of maximum dose) are performed with and without a parallel magnetic field (0.22T at magnet center).
Results:
In the presence of a magnetic field, the unobstructed surface dose is higher (31.9%Dmax versus 22.2%Dmax). The surface dose is found to increase linearly with thickness for thin (<25 μm) copper (0.339%Dmax μm-1) and aluminum (0.116%Dmax μm-1) foils. In the presence of a magnetic field the slope is lower (copper: 0.16%Dmax μm-1, aluminum: 0.06%Dmax μm-1). The effect of in-beam foils is reduced due to partial shielding of the surface from contaminant electrons. Copper causes a surface dose increase ≈3 times higher than aluminum of the same thickness, consistent with their relative electron density. Polyester film (127μm) increases the surface dose (to 35% Dmax with field) about as much as a gown (36% Dmax with field), while the increase with polyimide film (17μm) is less than 1% above the open field dose.
Conclusions:
Thin copper and aluminum conductors increase surface dose by an amount comparable to a hospital gown. Similarly, enclosure materials made of thin polyester or polyimide film increase surface dose by only a few %Dmax in excess of an unobstructed beam. Based on measurements in this study, in-beam, surface RF coils are feasible for MRIgRT systems.
Related Concept Videos
Magnetic Field Of A Current Loop
Magnetic Field Due to Two Straight Wires
Magnetic Fields
A magnetic field is defined by the force that a charged particle experiences...
Magnetic Force Between Two Parallel Currents
The force exerted by the magnetic field due to the first conductor over a finite length of the second conductor is given as the product of the current in the second conductor and the vector product of the length vector along the current element and the field due to the first conductor. According to the...
Magnetic Field Due To A Thin Straight Wire
Diamagnetic Shielding of Nuclei: Local Diamagnetic Current