Related Experiment Video
Updated: Aug 5, 2026

A Finite Element Approach for Locating the Center of Resistance of Maxillary Teeth
Published on: April 8, 2020
Design optimization of implant geometrical characteristics enhancing primary stability using FEA of stress
Sameh Elleuch1, Hanen Jrad1, Amir Kessentini1,2
1Laboratory of Electromechanical Systems (LASEM), National Engineering School of Sfax, University of Sfax, Sfax, Tunisia.
Abstract:
The main objective of this study was to investigate the influence of implant geometrical characteristics: diameter, length and thread's pitch, on stress distribution around dental prosthesis. A set of numerical simulations using FEM were conducted and responses surfaces were generated. With the aim of optimizing the equivalent stresses responses; desirability function approach was adopted to solve this multi-objective problem. Results showed that implant diameter had most significant influence on generated stresses and high concentration of stresses were identified in the lower part of the implant. This study is helpful in choosing the optimal dental implant for clinical application.