Detection of cardiac allograft vasculopathy by multi-layer left ventricular longitudinal strain in heart transplant

C Sciaccaluga1, G E Mandoli2, N Sisti2

  • 1Department of Medical Biotechnologies, Section of Cardiology, University of Siena, Siena, Italy. carlotta.sciaccaluga@gmail.com.

Insights

Speckle tracking echocardiography (STE) shows promise for early detection of cardiac allograft vasculopathy (CAV) in heart transplant recipients. The endocardial-epicardial strain gradient, along with diastolic parameters, are key non-invasive predictors of CAV.

Area of Science:

  • Cardiology
  • Transplant Medicine
  • Medical Imaging

Background:

  • Cardiac allograft vasculopathy (CAV) is a primary cause of heart transplant failure.
  • Diagnosing CAV non-invasively in early stages remains a significant clinical challenge.
  • Current gold standard diagnosis relies on invasive coronary angiography (CA).

Purpose of the Study:

  • To evaluate the efficacy of speckle tracking echocardiography (STE), specifically three-layer STE, in predicting early-stage CAV.
  • To identify other echocardiographic, clinical, or biochemical parameters associated with CAV.
  • To assess the diagnostic performance of novel echocardiographic markers for CAV.

Main Methods:

  • A cohort of 33 heart transplant patients were divided into CAV+ (12) and CAV- (22) groups.
  • All patients underwent transthoracic echocardiography and coronary angiography (CA) on the same day.
  • Speckle tracking echocardiography (STE) was used to assess myocardial strain values across different layers.

Main Results:

  • Reduced strain values were observed in all myocardial layers in patients with CAV.
  • The endocardial-epicardial longitudinal strain (LS) gradient demonstrated high predictive value for CAV (AUC 0.97).
  • E wave deceleration time (DT) and mean E/e' ratio were strongly associated with CAV presence.

Conclusions:

  • Layer-specific STE, particularly the endocardial-epicardial LS gradient, shows significant potential for non-invasive early detection of CAV.
  • E wave DT and E/e' ratio are valuable independent predictors of CAV.
  • STE offers a promising non-invasive tool to complement CA for CAV monitoring in heart transplant patients.