Image optimization and reduction of radiation dose in CT of the paranasal sinuses
Mohammad Davoudi1, Shaghayegh Lotfian Ahangarkolaei2
1Department of Medical Imaging Center, Babol University of Medical Sciences, Babol, Iran.
Introduction:
Due to use of ionization radiation in the computed tomography (CT), optimal parameters should be used to reduce the risk of incidence of secondary cancers in patients who are constantly exposed to x-rays. To reduce the dose delivered to patients in each scan, CT technologists can change the image acquisition parameters. However, this reduces image quality. The present study aimed to optimize and reduce radiation dose in a CT of the paranasal sinuses while minimizing deterioration of image quality.
Methods:
In this study patients were divided in two groups: Group A was scanned axially and coronally using default parameters, while Group B was scanned axially and coronally using new parameters. Common CT dose descriptors including weighted computed tomography dose index (CTDIw), volumetric CTDI (CTDIvol), dose length product (DLP), effective dose (ED) and image noise were measured for each group. The patients' organ doses were estimated using the ImPACT CT patient Dosimetry Calculator. The tube voltage, tube current, pitch, rotation time, and other parameters were then reduced and optimized. After reconstruction and analysis, all of the images were of good diagnostic quality in both groups Results: Using the new parameters, good agreement was found between the direct and reconstructed images. The CT parameters were reduced by the following proportions: kVp-16.6%, mA-75%, rotation time-20%, and mAs-80%. However, these reductions did not obscure any anatomical landmarks. These parameters reduced the CTDIw, CTDIvol and DLP by 88.2%, 91.3%, and 91.3% respectively.
Conclusion:
The results suggest that the use of a Bone algorithm reduces the total amount of radiation used during CT of the sinuses. We recommend using these parameters in children, in the evaluation of facial trauma, and in emergency CT of the paranasal sinuses.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...