Related Experiment Video
Updated: Nov 21, 2025

MRM Microcoil Performance Calibration and Usage Demonstrated on Medicago truncatula Roots at 22 T
Published on: January 16, 2021
Design and construction of a gradient coil for high resolution marmoset imaging
William B Handler1, Geron Bindseil2, Ryan Chaddock1
1xMR lab, Physics and Astronomy Department, Western University, London, Ontario, Canada.
None:
A gradient coil with integrated second and third order shims has been designed and constructed for use inside an actively shielded 310 mm horizontal bore 9.4 T small animal MRI. An extension of the boundary element method, to minimise the power deposited in conducting surfaces, was used to design the gradients, and a boundary element method with a constraint on mutual inductance was used to design the shims. The gradient coil allows for improved imaging performance and was optimized for an imaging region appropriate for marmoset imaging studies. Efficiencies of 1.5 mT m-1 A-1 were achieved in a 15 cm wide bore while maintaining gradient uniformity ≤5% over the 8 cm region of interest. Two new cooling methods were implemented which allowed the gradient coil to operate at 100 A RMS, 25 % of max current with a temperature rise below 30 C.