Published on: December 19, 2020
Implementing a Novel Through-Glass Chest Radiography Technique for COVID-19 Patients: Image Quality, Radiation Dose
Jessica Chan1, William Auffermann1, Peter Jenkins2
1University of Utah School of Medicine, Department of Radiology and Imaging Sciences, Salt Lake City, UT.
Rationale And Objectives:
The novel coronavirus (COVID-19) pandemic has presented many logistical challenges, including unprecedented shortages of personal protective equipment (PPE). A technique of obtaining portable chest radiographs (pCXR) through glass doors or windows to minimize technologist-patient contact and conserve PPE has gained popularity, but remains incompletely evaluated in the literature. Our goal was to quickly implement this technique and evaluate image quality and radiation dose.
Materials And Methods:
An infographic and video were developed to educate nurses and technologists on the through-glass pCXR technique. Imaging parameters were optimized using a phantom and scatter radiation was measured. Three reviewers independently evaluated 100 conventionally obtained and 100 through-glass pCXRs from March 13, 2020 to April 30, 2020 on patients with suspected COVID-19, using criteria for positioning and sharpness/contrast on a 1 (confident criteria not met) to 5 (confident criteria met) scale. Imaging parameters, including deviation index (DI) were recorded for all radiographs.
Results:
The through-glass method was rapidly adopted and conserved one isolation gown per interaction. Although there was a statistically significant difference in the positioning (P value 0.018) and sharpness/contrast (P value 0.016), the difference in mean ratings was small: 4.82 vs 4.65 for positioning and 4.67 vs 4.50 (conventional vs modified) for sharpness/contrast. Scatter radiation was measured using a thorax phantom and found to be acceptable for the patient and nearby personnel. Standard deviation was higher for the DI for the through-glass technique (2.8) compared to the conventional technique (1.8), although the means were similar.
Conclusion:
The through-glass technique was quickly implemented, producing diagnostic quality chest radiographs while conserving PPE and reducing risks to radiology staff. There was more variability with imaging technique and DI using the through-glass technique, likely due to technologist uncertainty regarding technical modifications. Further work to reduce this variation is necessary to optimize quality and dose.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT