First clinical experience with the novel cold storage SherpaPak™ system for donor heart transportation
Dejan Radakovic1, Seymur Karimli1, Kiril Penov1
1Department of Thoracic and Cardiovascular Surgery, University Hospital Würzburg, University of Wuerzburg, Würzburg, Germany.
Background:
The current gold standard for donor heart preservation is a three-bag-technique and storage in a cooler filled with slush ice. This technique can cause freezing injury with protein denaturation. We report our early experience with a single-use disposable device (SherpaPak™, Paragonix Technologies, MA, USA) specifically designed for sterile permanent temperature-controlled transportation of donor hearts.
Methods:
In this case control study with 2:1 matching we identified 21 patients after heart transplantation depending on type of organ transport (standard three-bag-technique vs. SherpaPak™). The outcome after donor heart storage in the SherpaPak™ was compared with donor heart transportation with the standard technique.
Results:
Since July 2018 seven patients (5 males; mean age 50.3±13.2years) underwent heart transplantation with the SherpaPak™ system. Cold ischemic time was longer in the SherpaPak™ group (207.7±23.3 vs. 181.6±21.9, P=0.027). SherpaPak™ kept the organ temperature at 5.1±0.8 °C, with an average outside temperature of 21.4±3.6 °C. Among all 21 transplanted patients four developed fatal early graft failure (28.6% vs. 21.4%, P=0.432). Over the first hours we noticed no difference in hemodynamic parameters, CK-MB levels or vasoactive-inotropic score. During first follow-up we noticed slightly better right heart function in the SherpaPak™ group (TAPSE 17.83±2.71 vs. 14.52±2.61 mm, P=0.020). We identified no positive blood cultures in the SherpaPak™ group within the first 30 days after heart transplantation.
Conclusions:
The SherpaPak™ provides a constant temperature during transportation with permanent monitoring, never dropping below 4 °C. Organs transported with this novel device showed a normal perioperative function.