Pressure infusor devices. Do they generate the pressures indicated?

P Cox1

  • 1Department of Anaesthesia, Motala Hospital, Sweden.

Anaesthesia
|January 1, 1988
PubMed
Summary

Pressure infusor devices show significant pressure discrepancies between the manometer reading and actual fluid pressure. These inaccuracies, dependent on remaining fluid volume, can lead to critical pressure variations during infusion.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Non-hormonal treatments for managing vulvovaginal atrophy/genitourinary syndrome of menopause.

Climacteric : the journal of the International Menopause Society·2023
Same author

Cause of intrauterine and neonatal death in twin pregnancies (CoDiT): development of a novel classification system.

BJOG : an international journal of obstetrics and gynaecology·2020
Same author

Vulvovaginal atrophy in women after cancer.

Climacteric : the journal of the International Menopause Society·2019
Same author

The first clonal spread of vanA-positive Enterococcus raffinosus in a nursing home.

The Journal of hospital infection·2017
Same author

Re: Stillbirth collection by Man et al.

Ultrasound in obstetrics & gynecology : the official journal of the International Society of Ultrasound in Obstetrics and Gynecology·2017
Same author

Efficient development of highly polymorphic microsatellite markers based on polymorphic repeats in transcriptome sequences of multiple individuals.

Molecular ecology resources·2014

Area of Science:

  • Medical Devices
  • Biomedical Engineering
  • Clinical Engineering

Background:

  • Pressure infusor devices are commonly used in healthcare settings to rapidly administer intravenous fluids.
  • Accurate pressure delivery is crucial for effective and safe fluid management, especially in critical care scenarios.

Purpose of the Study:

  • To evaluate the accuracy of pressure readings on three different pressure infusor devices.
  • To determine the correlation between the infusor manometer pressure and the actual pressure within the infusion system.

Main Methods:

  • Tested three distinct pressure infusor devices.
  • Measured infusion system pressure at a set infusor bag pressure (300 mmHg) while emptying fluid in 50 ml aliquots.
  • Determined the infusor bag pressure needed to maintain 300 mmHg as fluid was stepwise removed.

Main Results:

  • Significant discrepancies observed between infusor manometer readings and actual fluid pressure, unrelated to manometer error.
  • Pressure variations ranged from 170 mmHg above to 200 mmHg below the indicated 300 mmHg.
  • One device failed to maintain 300 mmHg pressure with 100 ml of fluid remaining.

Conclusions:

  • Current pressure infusor devices exhibit substantial inaccuracies in pressure delivery.
  • The amount of fluid remaining in the infusion bag significantly impacts pressure accuracy.
  • These findings highlight potential risks and necessitate careful consideration of device performance in clinical practice.