Inter-vendor variability in strain measurements depends on software rather than image characteristics
Serkan Ünlü1,2, Oana Mirea1,3, Stéphanie Bézy1
1Department of Cardiovascular Diseases, University Hospital Leuven, Herestraat 49, 3000, Leuven, Belgium.
Vendor-independent software for echocardiographic strain analysis shows consistent results across different ultrasound vendors. While tracking feasibility varied, strain measurement reproducibility and scar detection were largely unaffected by vendor-specific image characteristics.
Area of Science:
- Cardiology
- Medical Imaging
- Echocardiography
Background:
- Echocardiographic strain measurements exhibit high inter-vendor variability due to proprietary algorithms.
- Vendor-independent software offers a potential solution to standardize strain analysis.
- The influence of vendor-specific image characteristics on vendor-independent software tracking is not well understood.
Purpose of the Study:
- To investigate the reproducibility, accuracy, and scar detection ability of vendor-independent strain measurements.
- To assess how vendor-specific image characteristics impact strain analysis using vendor-independent software.
Main Methods:
- Utilized vendor-independent software (TomTec Image Arena) on 63 patient datasets from four ultrasound vendors (GE, Philips, Siemens, Toshiba).
- Evaluated tracking feasibility, inter-vendor bias, test-re-test variability, and scar discrimination ability.
- Used cardiac magnetic resonance delayed enhancement imaging as the reference standard for scar definition.
Main Results:
- Tracking feasibility differed significantly across vendors (p < 0.001).
- Variability and reproducibility of global longitudinal strain (GLS) and segmental longitudinal strain (SLS) were similar among vendors.
- Scar detection capability and regional dysfunction identification showed no significant differences across vendors.
Conclusions:
- Strain measurement results are primarily dependent on the analysis software, with minimal influence from vendor-specific image characteristics.
- Vendor-independent software provides reproducible strain measurements and reliable scar detection across different ultrasound platforms.
- Standardization of echocardiographic strain analysis can be improved by using vendor-independent software.