Leveraging Turbidity and Thromboelastography for Complementary Clot Characterization
Published on: June 4, 2020
Revisiting the effects of spectral interfering substances in optical end-point coagulation assays
Jansen N Seheult1,2, Daniel Dalenberg3, Meera R Sridharan3
1Vitalant Specialty Labs - Coag & MID, Pittsburgh, PA, USA.
Introduction:
Hemolysis, icterus, and lipemia (HIL) are common pre-analytical variables in the clinical laboratory. Understanding their effects on coagulation laboratory results is essential.
Methods:
HIL effects on the prothrombin time (PT), activated partial thromboplastin time (APTT), dilute Russell's viper venom time (DRVVT), thrombin time (TT), and protein C chromogenic activity (CFx) were evaluated on the ACL TOP 750 optical analyzer and STA-R Evolution mechanical analyzer (PT and APTT only) by spiking normal donor, patient, and commercial control samples with varying concentrations of hemolysate, bilirubin, or a lipid emulsion. The relative difference or bias compared to the original results was determined.
Results:
Hemolysis (H) indices up to 900 mg/dL did not affect the APTT, PT, DRVVT Confirm, TT, and CFx; however, H indices above approximately 200 mg/dL resulted in a false-negative DRVVT screen and screen/confirm ratio in samples with a lupus anticoagulant. There was an artifactual prolongation of the PT and APTT when conjugated bilirubin was dissolved in aqueous solvents and not when it was dissolved in dimethyl sulfoxide. Icterus (I) indices up to 45 mg/dL did not result in significant (>15%) bias for all assays evaluated. The PT and APTT assays failed to produce a robust clot curve when the lipemia (L) index exceeded 6000 milliabsorbance units (mAbs), and the TT and DRVVT assays failed when the L index exceeded 3000 mAbs; the CFx assay was unaffected by lipemia.
Conclusions:
Verification of the manufacturer's recommended interference thresholds is important since it may avoid inappropriate instrument flagging and/ or sample rejection.
Related Concept Videos
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Emission Spectroscopy: Interference
Spectrophotometry: Introduction
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...