Compressed sensing time-of-flight magnetic resonance angiography with high spatial resolution for evaluating intracranial aneurysms: comparison with digital subtraction angiography
Donghyun Kim1, Young Jin Heo1, Hae Woong Jeong1
1Department of Radiology, Inje University Busan Paik Hospital, Republic of Korea.
Background And Purpose:
Compressed sensing is used for accelerated acquisitions with incoherently under-sampled k-space data, and intracranial time-of-flight magnetic resonance angiography is suitable for compressed sensing. Compressed sensing time-of-flight is beneficial in decreasing acquisition time and increasing spatial resolution while maintaining acquisition time. In this retrospective study, we aimed to evaluate the image quality and diagnostic performance of compressed sensing time-of-flight with high spatial resolution and compare with parallel imaging time-of-flight using digital subtraction angiography as a reference.
Material And Methods:
In total, 39 patients with 46 intracranial aneurysms underwent parallel imaging and compressed sensing time-of-flight in the same imaging session and digital subtraction angiography before or after magnetic resonance angiography. The overall image quality, artefacts and diagnostic confidence were assessed by two observers. The contrast ratio, maximal aneurysm diameters and diagnostic performance were evaluated.
Results:
Compressed sensing time-of-flight showed significantly better overall image quality, degree of artefacts and diagnostic confidence in both observers, with better inter-observer agreement. The contrast ratio was significantly higher for compressed sensing time-of-flight than for parallel imaging time-of-flight in both observers (source images, P < 0.001; maximum intensity projection images, P < 0.05 for both observers); however, the measured maximal diameters of aneurysms were not significantly different. Compressed sensing time-of-flight showed higher sensitivity, specificity, accuracy and positive and negative predictive values for detecting aneurysms than parallel imaging time-of-flight in both observers, with better inter-observer agreement. Compressed sensing time-of-flight was preferred over parallel imaging time-of-flight by both observers; however, parallel imaging time-of-flight was preferred in cases of giant and large aneurysms.
Conclusions:
Compressed sensing-time-of-flight provides better image quality and diagnostic performance than parallel imaging time-of-flight. However, neuroradiologists should be aware of under-sampling artefacts caused by compressed sensing.
More Related Videos
10:41Analysis of Cerebral Vasospasm in a Murine Model of Subarachnoid Hemorrhage with High Frequency Transcranial Duplex Ultrasound
Published on: June 3, 2021
09:36A Magnetic Resonance Imaging-based Computational Protocol for Analysis of Plaque Morphology and Hemodynamics in Patients with Carotid Artery Stenosis
Published on: August 12, 2025
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI