Systematic analysis and comparison of commercial seizure-detection software
Johannes Koren1,2, Sebastian Hafner2, Moritz Feigl3,4
1Karl Landsteiner Institute for Clinical Epilepsy Research and Cognitive Neurology, Vienna, Austria.
Epilepsia
|January 19, 2021
Summary
Three seizure-detection software packages demonstrated comparable seizure detection rates in epilepsy patients. However, differences in false alarm rates and detection delays were observed, with Persyst 13 showing the shortest delay and Encevis 1.7 the lowest false alarm rate.
Area of Science:
- Neurology
- Medical Technology
- Epilepsy Research
Background:
- Epilepsy management relies on accurate seizure detection.
- Long-term video-electroencephalography monitoring (VEM) is crucial for diagnosis and treatment.
- Automated seizure detection software aims to improve efficiency and accuracy.
Purpose of the Study:
- To evaluate the performance of three commercial seizure-detection software packages: Besa 2.0, Encevis 1.7, and Persyst 13.
- To compare their sensitivity, specificity (false alarm rate), and detection delay in epilepsy patients undergoing VEM.
- To assess the accuracy of these tools in a clinical setting.
Main Methods:
- A comparative study involving 81 epilepsy patients undergoing long-term VEM.
- Analysis of seizure detection performance metrics: sensitivity, specificity, and detection delay.
- Statistical comparison of Besa 2.0, Encevis 1.7, and Persyst 13 using kappa statistics.
Main Results:
- No significant difference in per-patient seizure detection rates (sensitivity) among Besa (67.6%), Encevis (77.8%), and Persyst (81%).
- Significant differences in false alarm rates (per hour): Besa (0.7/h), Encevis (0.2/h), Persyst (0.9/h).
- Significant differences in detection delay (seconds): Besa (30s), Encevis (25s), Persyst (20s). Persyst 13 had the shortest delay.
Conclusions:
- Commercially available seizure-detection software packages offer comparable sensitivity for seizure detection.
- Significant variations exist in false alarm rates and detection delays among these software packages.
- Persyst 13 exhibited the highest detection rate and shortest delay, while Encevis 1.7 had the lowest false alarm rate.