The Accuracy of Smartphone Sound Level Meter Applications (SLMAs) in Measuring Sound Levels in Clinical Rooms
Yula C Serpanos1,2, Janet R Schoepflin1,2, Steven R Cox1
1Department of Communication Sciences and Disorders, Adelphi University, Garden City, New York.
Calibrating smartphone sound level meter applications (SLMAs) with external or internal microphones improves accuracy for measuring sound levels above 40 dBA. However, SLMAs may overestimate lower sound levels, necessitating careful use in clinical settings.
Area of Science:
- Audiology
- Acoustics
- Mobile Technology
Background:
- Smartphone sound level meter applications (SLMAs) show variable accuracy due to device and app differences.
- Previous research indicates inconsistent performance of SLMAs in measuring sound levels.
Purpose of the Study:
- To assess the accuracy of smartphone SLMAs for measuring sound levels in clinical rooms.
- To compare SLMA accuracy with and without calibration of internal and external microphones.
Main Methods:
- A quasi-experimental design was employed using two iOS SLMAs on an iPhone 6S.
- Measurements included white noise stimuli and sound levels in quiet and simulated background noise.
- Accuracy was compared against a Type 1 sound level meter using chi-square analysis.
Main Results:
- Calibrated SLMAs demonstrated significantly higher accuracy for sound levels ≥ 40 dBA.
- Both calibrated and uncalibrated SLMAs overestimated sound levels below 40 dBA.
- SLMAs with calibrated microphones can verify audiological screening environments.
Conclusions:
- Calibrating smartphone microphones enhances SLMA accuracy for clinical sound level measurements above 40 dBA.
- Clinicians should calibrate microphones before using smartphone SLMAs for room and background noise monitoring.
- Overestimation of low-level sounds (< 40 dBA) by SLMAs requires attention during clinical use.
Related Concept Videos
Sound Intensity Level
The human ear can perceive an extensive range of sound intensity, necessitating the use of the logarithmic scale to define a physical quantity—the intensity level. It is a ratio of two intensities and...
Applications of Logarithms
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.