The impact of visual axis position on the optical quality after implantation of multifocal intraocular lenses with different asphericity values
Related Experiment Video
Updated: Nov 20, 2025

Author Spotlight: Advancements in Refractive Surgical Correction for Presbyopia and Exploring Postoperative Visual Acuity
Published on: September 20, 2024
Kata Miháltz1,2, Pia Veronika Vécsei-Marlovits3,4
1Department of Ophthalmology, Hospital Hietzing, Wolkersbergenstrasse 1, 1130, Vienna, Austria. kata.mihaltz@wienkav.at.
Purpose:
To evaluate the effect of visual axis positioning on the optical performance of the Tecnis MIOL and the Diff-aA MIOL.
Methods:
In this prospective, randomized comparative study, 70 eyes of 35 subjects with senile cataract were implanted with the spherical aberration-correcting diffractive, bifocal Tecnis ZLB00 IOL and 60 eyes of 30 age-matched subjects with the spherical aberration neutral, diffractive, bifocal Diffractiva IOL. Observation procedure was performed 1, 3, and 6 months postoperatively. Main outcome measures included uncorrected and corrected distance and near visual acuity, manifest refraction, ocular aberrations, and visual quality metrics with 2 mm and 4 mm pupil and the position of visual axis.
Results:
At the 6-month visit, no significant difference was found in monocular and binocular uncorrected (UDVA) and corrected (CDVA) distance and near (UNVA, CNVA) visual acuity between the groups. Spherical and coma-like aberrations were similar measured with a 2-mm pupil, but with a 4-mm pupil, the SA was significantly larger (in negative direction) in the Diffractiva group. The higher-order Strehl ratio and MTF was significantly larger in the Diffractiva group measured at 2 mm entrance pupil; however, this difference disappeared by the 4-mm pupil measurements. Postoperative angle alpha distance had a significant influence on HO Strehl value.
Conclusions:
The size of angle alpha is a predictive factor of image quality by multifocal IOL patients.
Trial Registration:
Trial registration number and date of registration: NCT04274088, 14.02.2020.
More Related Videos
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
06:55Scanning Light Scattering Profiler SLPS Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
Related Concept Videos
Focusing of Light in the Eye
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Glaucoma: Overview