Transcatheter Aortic Valve Replacement With the LOTUS Edge System: Early European Experience
Xavier Armario1, Liesbeth Rosseel1, Rajesh Kharbanda2
1Department of Cardiology, Galway University Hospital, Galway, Ireland.
JACC. Cardiovascular Interventions
|January 22, 2021
Summary
Transcatheter aortic valve replacement (TAVR) with the LOTUS Edge system shows good short-term safety and effectiveness, particularly in complex patients. However, the need for permanent pacemaker implantation remains high.
Area of Science:
- Cardiology
- Interventional Cardiology
- Medical Devices
Background:
- The LOTUS Edge system, a transcatheter aortic valve replacement device, was re-released in April 2019.
- This study reports the first European clinical experience with the LOTUS Edge system and its implantation technique.
Purpose of the Study:
- To evaluate the short-term safety and efficacy of transcatheter aortic valve replacement (TAVR) using the LOTUS Edge system.
- To assess clinical outcomes, including major adverse events and hemodynamic performance, in patients receiving the LOTUS Edge system.
Main Methods:
- A multicenter, single-arm, retrospective registry study.
- Inclusion of 286 consecutive patients undergoing TAVR with the LOTUS Edge system across 18 European centers.
- Analysis of clinical, echocardiographic, and computed tomographic data, with endpoints defined by Valve Academic Research Consortium-2 criteria.
Main Results:
- The study included 286 patients with a mean age of 81.2 years; 47.9% had complex anatomy.
- Thirty-day mortality was 2.4% and stroke rate was 3.5%.
- Excellent hemodynamic results were observed, with no severe paravalvular leak (PVL); however, new permanent pacemaker (PPM) implantation was required in 25.9% of patients.
Conclusions:
- The LOTUS Edge system demonstrated satisfactory short-term safety and efficacy in a complex patient cohort.
- Favorable hemodynamic results and very low rates of paravalvular leak were noted.
- High rates of permanent pacemaker implantation warrant further investigation into operator experience and technique optimization.