Imaging-Based Localization of His Bundle Pacing Electrodes: Results From the Prospective IMAGE-HBP Study
Pugazhendhi Vijayaraman1, Gopi Dandamudi2, Faiz A Subzposh1
1Geisinger Heart Institute, Wilkes-Barre, Pennsylvania, USA.
Objectives:
This study sought to evaluate the correlation between His bundle (HB) pacing (HBP) implantation characteristics, lead-tip location, and association of intraprocedural His recordings with approximated HB anatomic landmarks using computed tomography (CT) imaging.
Background:
HBP continues to grow in clinical practice due to offering true physiological pacing. However, a clear understanding of HB anatomy and the lead-tip location's influence on pacing characteristics is lacking.
Methods:
The IMAGE-HBP study (Imaging Study of Lead Implant for His Bundle Pacing) was a prospective, multicenter study designed to assess implantation characteristics of the SelectSecure Model 3830 lead placed at the HB, evaluate protocol-specified HBP success (His recording present on electrogram and HBP threshold ≤2.5 V at 1 ms), and correlation between lead-tip location by CT imaging and HBP characteristics as well as lead-related complications through 12 months.
Results:
Sixty-nine patients underwent a lead implantation attempt at the HB. Of these, 61 patients (88%) had a lead successfully implanted at the HB, and 52 patients (75%) met the pre-specified definition of successful HBP. In 51 patients with CT imaging, 11 leads (22%) were placed in the atrial aspect of the HB region (36% selective HBP), and 40 leads (78%) were placed in the ventricular aspect (28% selective HBP). Four of the 51 patients had P-wave oversensing, all with leads in the atrium. Freedom from lead-related complication at 12 months was 93%.
Conclusions:
Successful HBP could be achieved at lead-tip locations in the atrium or ventricle but is preferable in the ventricle to eliminate risk of oversensing. The IMAGE-HBP study offers better insight into approximated HB anatomic landmarks, lead-tip location, and correlation with pacing characteristics. (Imaging Study of Lead Implant for His Bundle Pacing [IMAGE-HBP]; NCT03294317).