Fit-failure rate associated with simulated reuse and extended use of N95 respirators assessed by a quantitative fit
Jiwon Jung1,2, Jiyun Kim2, Hyejin Yang2
1Department of Infectious Diseases, Asan Medical Center, University of Ulsan College of Medicine, Seoul, South Korea.
Objective:
We quantitatively assessed the fit failure rate of N95 respirators according to the number of donning/doffing and hours worn.
Design:
Experimental study.
Setting:
A tertiary-care referral center in South Korea.
Participants:
In total, 10 infection control practitioners participated in the fit test.
Methods:
The first experiment comprised 4 consecutive 1-hour donnings and fit tests between each donning. The second experiment comprised 2 consecutive 3-hour donnings and fit tests between each donning. The final experiment comprised fit tests after an 1-hour donning or a 2-hour donning.
Results:
For 1-hour donnings, 60%, 70%, and 90% of the participants had fit failures after 2, 3, and 4 consecutive donnings, respectively. For 3-hour donnings, 50% had fit failure after the first donning and 70% had failures after 2 consecutive donnings. All participants passed the fit test after refitting whenever fit failure occurred. The final experiment showed that 50% had fit failure after a single use of 1 hour, and 30% had fit failure after a single use of 2 hours.
Conclusions:
High fit-failure rates were recorded after repeated donning and extended use of N95 respirators. Caution is needed for reuse (≥1 time) and extended use (≥1 hour) of N95 respirators in high-risk settings such as those involving aerosol-generating procedures. Although adequate refitting may recover the fit factor, the use of clean gloves and strict hand hygiene afterward should be ensured when touching the outer surfaces of N95 respirators for refitting.
More Related Videos
07:30A Test Bed to Examine Helmet Fit and Retention and Biomechanical Measures of Head and Neck Injury in Simulated Impact
Published on: September 21, 2017
08:44Author Spotlight: Integrating Alveolar-Capillary Reserve Measurements in Exercise Adaptation and Therapeutic Strategies
Published on: February 2, 2024
Related Concept Videos
Expected Frequencies in Goodness-of-Fit Tests
Fatigue
Goodness-of-Fit Test
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: